DMM measurement tolerances
Statement of accuracy in ± (% of reading + display error in counts (= number of smallest in-
dicated point)). The accuracy is valid for one year at a temperature of +23°C ± 5°C, and at a
relative humidity of less than 75 %, non-condensing.
Direct voltage range, overload protection 600 V
Range Accuracy Resolution
200 mV ±(0,5% + 3) 0,1 mV
2 V
±(1% + 3)
1 mV
20 V 10 mV
200 V 100 mV
600 V 1 V
Alternate voltage range, overload protection 600 V
Area (50-60 Hz) Accuracy Resolution
2 V
±(1% + 5)
1 mV
20 V 10 mV
200 V
±(1,5% + 10)
100 mV
600 V 1 V
Resistance range
Range Accuracy Resolution
200 Ω ±(0,8% + 5) 0.1 Ω
2 kΩ
±(1,2% + 3)
1 Ω
20 kΩ 10 Ω
200 kΩ 100 Ω
2 MΩ ±(2% + 5) 1 kΩ
20 MΩ ±(5% + 8) 10 kΩ
Acoustic Continuity tester <150 Ω
Diode test voltage 1.5V/DC; test current: max. 1 mA