Instructions
Table Of Contents
- 1.General Safety Requirements
- 2.Safety Terms and Symbols
- 3.Junior User Guidebook
- Introduction to the Structure of the Oscilloscope
- User Interface Introduction
- How to Implement the General Inspection
- How to Implement the Function Inspection
- How to Implement the Probe Compensation
- How to Set the Probe Attenuation Coefficient
- How to Use the Probe Safely
- How to Implement Self-calibration
- Introduction to the Vertical System
- Introduction to the Horizontal System
- Introduction to the Trigger System
- Touchscreen Controls
- Magnifier Function
- 4.Advanced User Guidebook
- How to Set the Vertical System
- Use Mathematical Manipulation Function
- Use Vertical Position and Scale Knobs
- How to Set the Horizontal System
- How to Set the Trigger/Decoding System
- How to Operate the Function Menu
- How to Implement Sampling Setup
- How to Set the Display System
- How to Save and Recall a Waveform
- How to Record/Playback Waveforms
- How to Clone and Recall a waveform
- How to Implement the Auxiliary System Function Set
- How to Update your Instrument Firmware
- How to Measure Automatically
- How to Measure with Cursors
- How to Use Autoscale
- How to Use Built-in Help
- How to Use Executive Buttons
- How to Print the Screen Image
- 5.Use the Arbitrary Function Generator (Optional)
- 6.Use the Multimeter (Optional)
- 7.Communication with PC
- 8.Demonstration
- 9.Troubleshooting
- 10.Technical Specifications
- 11.Appendix
5.Use the Arbitrary Function Generator (Optional)
102
Gain Offset
Adjust offset of gain plot, range from -250dB to
250dB.
Phase Scale
Adjust scale of phase plot, range from 5° to 180°.
Phase Offset
Adjust offset of phase plot, range from -180° to 180°.
Autoscale
Autoscale gain and phase plots.
Analysis
Run/stop the analysis.
F: Frequency; G: Gain; P: Phase.
To run the frequency response analysis, do the following:
1. The waveform generator output is connected to a device under test. The input to the
device is probed by CH1of the oscilloscope. The output from the device is probed by
CH2of the oscilloscope.
2. Push the Utility button, select Function in the bottom menu, select FRA in the left
menu
3. In the bottom menu, select
FRA. In the right menu, check
FRA, and set other
menu items.
4. In the bottom menu, select Setup or Chart. In the right menu, set the menu items.
5. In the bottom menu, check Analysis to run the analysis.