Instructions
Table Of Contents
- 1.General Safety Requirements
- 2.Safety Terms and Symbols
- 3.Junior User Guidebook
- Introduction to the Structure of the Oscilloscope
- User Interface Introduction
- How to Implement the General Inspection
- How to Implement the Function Inspection
- How to Implement the Probe Compensation
- How to Set the Probe Attenuation Coefficient
- How to Use the Probe Safely
- How to Implement Self-calibration
- Introduction to the Vertical System
- Introduction to the Horizontal System
- Introduction to the Trigger System
- Touchscreen Controls
- Magnifier Function
- 4.Advanced User Guidebook
- How to Set the Vertical System
- Use Mathematical Manipulation Function
- Use Vertical Position and Scale Knobs
- How to Set the Horizontal System
- How to Set the Trigger/Decoding System
- How to Operate the Function Menu
- How to Implement Sampling Setup
- How to Set the Display System
- How to Save and Recall a Waveform
- How to Record/Playback Waveforms
- How to Clone and Recall a waveform
- How to Implement the Auxiliary System Function Set
- How to Update your Instrument Firmware
- How to Measure Automatically
- How to Measure with Cursors
- How to Use Autoscale
- How to Use Built-in Help
- How to Use Executive Buttons
- How to Print the Screen Image
- 5.Use the Arbitrary Function Generator (Optional)
- 6.Use the Multimeter (Optional)
- 7.Communication with PC
- 8.Demonstration
- 9.Troubleshooting
- 10.Technical Specifications
- 11.Appendix
7.Communication with PC
114
set the IP address to 192.168.1.71, Subnet mask is 255.255.255.0, Default gateway is
192.168.1.1.
Figure 7- 5 Set the network parameters of the computer
(3) Set the network parameters of the Oscilloscope Software. Run the software on the
computer; choose the "Ports-settings" of the "Communications" menu item. Set
"Connect using" to LAN. About the IP, the first three bytes is same as the IP in the
step (2), the last byte should be different. Here, we set it to 192.168.1.72. The range of
the port value is 0 - 4000, but the port which under 2000 is always used, so it is
suggested to set it to the value above 2000. Here, we set it to 3000.
Figure 7- 6 Set the network parameters of the Oscilloscope Software