Instructions

10.Technical Specifications
126
Performance Characteristics
Instruction
Bandwidth limit
20 MHz, full bandwidth
Channel to channel
isolation
50Hz: 100 : 1
10MHz: 40 : 1
Time delay between
channel(typical)
150ps
Sampling rate range
Dual
CH
8 bits mode
0.05 S/s
1 GS/s
12 bits mode
0.05 S/s
500 MS/s
14 bits mode
0.05 S/s
100 MS/s
Single
CH
8 bits mode
0.05 S/s
1 GS/s
12 bits mode
0.05 S/s
500 MS/s
14 bits mode
0.05 S/s
100 MS/s
Interpolation (Sinx)/x, x
Max Record length 40M
Scanning speed (S/div)
1ns/div - 1000s/div,
step by 1 2 - 5
Sampling rate / relay
time accuracy
±1 ppm (Typical, Ta = +25)
Interval(T) accuracy
(DC - 100MHz)
Single
±(1 interval time+1 ppm×reading+0.6 ns);
Average>16
±
(1 interval time +1 ppm×reading+0.4 ns)
Vertical system
Sensitivity
1 mV/div10 V/div
Displacement
±2 V (1 mV/div 50 mV/div)
±20 V (100 mV/div – 1 V/div)
±200 V (2 V/div 10 V/div)
Analog bandwidth
200 MHz
Single bandwidth
Full bandwidth
Low Frequency
10 Hz (at input, AC coupling, -3 dB)
Rise time
(at input, Typical)
8 bits mode
≤ 1.75 ns
12 bits mode
2.33 ns
14 bits mode
17.5 ns
DC gain accuracy
1 mV
3%
2 mV
2%
≥ 5 mV
1.5%
DC accuracy (average)
Delta Volts between any two averages of
≥16 waveforms acquired with the same
scope setup and ambient conditions (V):
±(3% rdg + 0.05 div)
Waveform inverted ON/OFF