Instructions
5.Use the Arbitrary Function Generator (Optional)
94
5. Use the Arbitrary Function Generator (Optional)
The oscilloscope contains an integrated dual-channel arbitrary function generator (25 MHz
or 50 MHz). The function generator provides 4 basic waveforms (sine, square, ramp, and
pulse) and 46 built-in arbitrary waveforms (Noise, Exponential rise, Exponential fall,
Sin(x)/x, Staircase, etc.). You can create a user-definable waveform and save it to internal
storage or USB device.
Output Connection
Push the Utility button, select Function in the bottom menu, select Output in the left
menu. In the bottom menu, select Output, in the right menu, select AG Output.
Connect the BNC cable to the port marked Out 1or Out 2 in the back of the oscilloscope.
Out 1 is the output of CH1, Out 2 is the output of CH2, also can be used as the port of
trigger signal output & Pass/Fail output.
To see the output of the generator, connect the other end of the BNC cable to one of the
input channels on the front of the oscilloscope.
To Set Channels
To Switch Channels in Menu Settings
Push button to switch between CH1 menu, CH2 menu, and Channel Copy menu.
To Turn On/Off Output of Channels
Push or to turn on/off output of the corresponding channel. The indicator
will be lighted when the corresponding channel is tuned on.
Channel Copy menu
Push button to switch to Channel Copy menu.
Copy Channel
Select CH2 To CH1 in the bottom menu to copy parameters of CH2 to CH1.
Select CH1 To CH2 in the bottom menu to copy parameters of CH1 to CH2.
Frequency Lock
Select Freq Lock in the bottom menu as On, the frequency of the two channels can be
adjusted synchronously.