Operating manual EN PTM-100 Universal thermometer BNC connection Waterproof Precise and fast
Table of contents Table of contents 1 Legal address of the manufacturer................................................................................................... 4 2 2.1 2.2 2.3 2.4 2.5 About this documentation ................................................................................................................. Foreword...............................................................................................................................................
Table of contents 8.3.1 Certificates.......................................................................................................................................... 21 9 Error and system messages............................................................................................................ 23 10 Disposal............................................................................................................................................. 24 11 Technical data..............
1 | Legal address of the manufacturer PTM-100 1 Legal address of the manufacturer Conrad Electronic SE Klaus-Conrad-Str. 1 D-92240 Hirschau http://www.conrad.com WEEE reg. no. DE 28001718 4 / 26 B-H86.0.02A.DB214-1.
PTM-100 About this documentation | 2 2 About this documentation 2.1 Foreword Read this document carefully and familiarise yourself with the operation of the product before you use it. Keep this document ready to hand and in the immediate vicinity of the product so that it is available to the personnel/user for reference at all times in case of doubt. The product was developed according to the state of the art and fulfils the requirements of the applicable European and national Directives.
2 | About this documentation PTM-100 Outcome of an action Result, consequence or effect of an instruction. Emphases In order to simplify legibility and provide a clearer overview, various sections / information are emphasised. – 1234 Display elements – Mechanical controls – Product functions – Product labels – Cross-reference [} p. 5] – Foot notes 2.5 Further information Software version of the product: – V1.
PTM-100 Safety | 3 3 Safety 3.1 Explanation of safety symbols DANGER This symbol warns of imminent danger which can result in death, severe bodily injury, or severe property damage in case of non-observance. CAUTION This symbol warns of potential dangers or harmful situations which can cause damage to the device or to the environment in case of non-observance.
3 | Safety PTM-100 CAUTION Erratic behaviour! On suspicion that the product can no longer be operated without danger, it must be decommissioned and prevented from recommissioning with appropriate labelling. The safety of the user can be impaired by the device if, for example, if it shows visible damage, it no longer works as specified or if it was stored for an extended period of time under unsuitable conditions.
PTM-100 Safety | 3 See Technical data [} p. 25]. 3.5 Qualified personnel For commissioning, operation and maintenance, the relevant personnel must have adequate knowledge of the measuring process and use of the measurements, for which purpose this document makes a valuable contribution. The instructions in this document must be understood, observed and followed.
4 | Description PTM-100 4 Description 4.1 Scope of delivery Please check to ensure the completeness of the product after opening the package. You should find the following components: – Quick reference guide – Handheld measuring device, ready for operation, including batteries 4.2 Job description The product offers precision, speed and reliability in a compact, ergonomic housing.
PTM-100 The product at a glance | 5 5 The product at a glance 5.1 The PTM-100 LCD Display PTM-100 BNC connection 5.2 Display elements Display Battery indicator Evaluation of the battery status Unit display Display of units, if applicable, with unstable symbol or type of mode, min/max/hold Main display Measurement of the current temperature or value for min/max/hold Auxiliary display Measurement of the current temperature in min/ max/hold mode with unit 5.
5 | The product at a glance PTM-100 Function key Press briefly Long press, 2s Freeze measurement Return to measurement display Call up next parameter Open menu, frozen measurement is displayed Close menu, changes are saved 5.4 Connections BNC connection Connection for temperature sensor Un/locking with rotating ring on the cable plug CAUTION Waterproofness! Waterproofness is only guaranteed for plug connections in the plugged-in state in combination with waterproof cable plugs.
PTM-100 Operation | 6 6 Operation 6.1 Commissioning 6.1.1 Explanation The product is switched on with the On/Off button. It may be necessary to configure the product after switching on. See Configuration [} p. 13]. Description Prerequisite – Sufficiently full batteries are inserted in the product. Instruction – Press On/Off button. Outcome of an action Information about the configuration of the product appears in the display. POFF T.OF T.SL Automatic shutoff Automatic shut-off activated.
6 | Operation PTM-100 Representation Call up menu Next parameter Change value 2s Press: Single step Save changes Discard changes 2s 2s Hold: Rapid change Outcome of an action The Configuration menu is closed after the last parameter. NOTE If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product. 6.2.
PTM-100 Operation | 6 Backlighting L,TE OFF Backlighting deactivated 15 30 60 120 240 Automatic shut-off of the backlighting after a selected time in seconds, during which no buttons have been pressed No automatic shut-off of the backlighting ON Temperature unit UN,T °( Temperature display in °C °F Temperature display in °F Factory settings IN,T Outcome of an action NO Use current configuration YES Reset product to factory settings.
6 | Operation PTM-100 5. In order to save the new parameter value, press and hold the Function key for longer than 1 second. Representation Call up menu Hold Outcome of an action Release The Configuration menu is closed after the last parameter. NOTE If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product. 6.2.
PTM-100 Operation | 6 – Display in water bath 36.6 °C – Display in water bath setpoint T.SL = 37.0 °C – T.OF = display zero point correction – setpoint zero point – T.OF = -0.2 °C – 0.0 °C = -0.2 °C – T.SL = (setpoint gradient correction / (display gradient correction – T.OF) – 1) *100 – T.SL = (37.0 °C / (36.6 °C – (-0.2)) -1) *100 = 0.54 Outcome of an action The changed value is saved and the Configuration menu is closed.
7 | Bases for measurement PTM-100 7 Bases for measurement 7.1 Sensor/device accuracy The product can be equipped with different interchangeable sensors. Temperature sensors are categorise in the following classes. EN 60751 Class B A AA = 1/3 DIN B Deviation +/- 0.3 °C +/- 0.5 % of measurement +/- 0.15 °C +/- 0.2 % of measurement +/- 0.1 °C +/- 0.
PTM-100 Bases for measurement | 7 7.2.5 Limit values Temperature range CAUTION Destruction of the measuring sensor! When conducting measurements in media with high or very low temperatures, there is a risk that the measuring sensor is not designed for such extremes. – The limit values must be observed! B-H86.0.02A.DB214-1.
8 | Maintenance PTM-100 8 Maintenance 8.1 Operating and maintenance notices NOTE The product and temperature probe must be handled with care and used in accordance with the technical data. Do not throw or strike. NOTE Plugs and sockets must be protected from soiling. NOTE If the product is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoided as a result. 8.2 Battery 8.2.
PTM-100 Maintenance | 8 NOTE Unnecessary screwing places the water-tightness of the product, among other things, at risk and should be avoided. NOTE Read the following handling instructions before replacing batteries and follow them step by step. If disregarded, the product could be damaged or the protection from moisture could be diminished. Description Proceed as follows to replace the batteries. Prerequisites – The product is switched off. – A suitable PH1 is available Instruction 1.
8 | Maintenance PTM-100 laboratories which have demonstrated their expertise in accordance with DIN EN ISO/ IEC 17025. The ISO calibration includes any necessary adjustment with the purpose of minimising a deviation of the measuring device. DAkkS calibration certificates are accompanied with a list of individual measurements before and after the adjustment, documentation and, if applicable, graphic representation, calculation of the expanded measuring uncertainty and traceability to the national standard.
PTM-100 Error and system messages | 9 9 Error and system messages Display ---- Meaning Possible causes Remedy No suitable measur- Incorrect measuring Connect a suitable measuring probe connected probe ing probe Measurement far Measuring probe or Measurement leaves the outside of the meas- product defect permissible range uring range Send in for repair Battery depleted Battery depleted Replace battery No display, unclear char- System error Error in the product Send in for repair acters or no Product is d
10 | Disposal PTM-100 10 Disposal NOTE The device must not be disposed of with household waste. If the product is disposed of, please take it to a municipal collection point, where it will be transported to a disposal company in accordance with requirements of hazardous goods laws. Otherwise, return it to us, freight prepaid. We will then arrange for the proper and environmentally-friendly disposal. Please dispose of empty batteries at the collection points intended for this purpose. 24 / 26 B-H86.0.
PTM-100 Technical data | 11 11 Technical data Measuring range temperature Accuracy temperature -200.0 +450.0 °C (-328.0 .. +842.0 °F) – Observe the permissible range of application of the sensor that is used! -20 .. +100 °C: ±0.1 K ±1 digit otherwise: ±0.2 % of measured value ± 2 digits Response time T90 water (0.
12 | Service PTM-100 12 Service 12.1 Manufacturer If you have any questions, please do not hesitate to contact us: Contact VOLTCRAFT Distributed by Conrad Electronic SE Klaus-Conrad-Str. 1 92240 Hirschau, Germany Tel.: +49 9604 40 87 87 Fax: +49 180 5 312110 kundenservice@conrad.de WEEE reg. no. DE 28001718 26 / 26 B-H86.0.02A.DB214-1.