Operating manual EN KBM-110 pH Tester BNC connection Waterproof Precise and fast
Table of contents Table of contents 1 Legal address of the manufacturer................................................................................................... 4 2 2.1 2.2 2.3 2.4 2.5 About this documentation ................................................................................................................. 5 Foreword...............................................................................................................................................
Table of contents 8.2.2 8.2.3 8.2.4 8.2.5 Opening the configuration menu......................................................................................................... 22 Configuring parameters of the configuration menu............................................................................. 23 Adjustment of the measuring input ..................................................................................................... 25 Configuring parameters of the adjustment menu.............
1 | Legal address of the manufacturer KBM-110 1 Legal address of the manufacturer Conrad Electronic SE Klaus-Conrad-Str. 1 D-92240 Hirschau http://www.conrad.com WEEE reg. no. DE 28001718 4 / 32 B-H86.0.12.DB214-1.
KBM-110 About this documentation | 2 2 About this documentation 2.1 Foreword Read this document carefully and familiarise yourself with the operation of the product before you use it. Keep this document ready to hand and in the immediate vicinity of the product so that it is available to the personnel/user for reference at all times in case of doubt. The product was developed according to the state of the art and fulfils the requirements of the applicable European and national Directives.
2 | About this documentation KBM-110 Outcome of an action Result, consequence or effect of an instruction. Emphases In order to simplify legibility and provide a clearer overview, various sections / information are emphasised. – 1234 Display elements – Mechanical controls – Product functions – Product labels – Cross-reference [} p. 5] – Foot notes 2.5 Further information Software version of the product: – V1.
KBM-110 Safety | 3 3 Safety 3.1 Explanation of safety symbols DANGER This symbol warns of imminent danger which can result in death, severe bodily injury, or severe property damage in case of non-observance. DANGER This symbol indicates danger for living tissue as well as a variety of materials, which can be damaged or destroyed when coming into contact with this chemical.
3 | Safety KBM-110 DANGER Incorrect area of application! In order to prevent erratic behaviour of the product, personal injury or property damage, the product must be used exclusively as described in the chapter Description [} p. 10] in the operating manual.
KBM-110 Safety | 3 NOTE If the product is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoided as a result. NOTE This product does not belong in children's hands! For this purpose, also refer to 2 Technical data [} 31] 3.4 Intended use The product is designed for measuring the pH value and Redox by means of suitable electrodes in water an aqueous media.
4 | Description KBM-110 4 Description 4.1 Scope of delivery Please check to ensure the completeness of the product after opening the package. You should find the following components: – Quick reference guide – Handheld measuring device, ready for operation, including batteries – Electrode GE 114 WD 4.2 Job description The product offers precision, speed and reliability in a compact, ergonomic housing.
KBM-110 The product at a glance | 5 5 The product at a glance 5.1 The KBM-110 LCD Display KBM-110 BNC connection and 2x4 mm banana KBM-110 5.2 Display elements Display Battery indicator Evaluation of the battery status Unit display Display of units, if applicable, with unstable symbol or type of mode, min/max/hold Main display Measurement of the current pH value or value for min/max/hold Auxiliary display Corresponding temperature for the displayed pH value with unit.
5 | The product at a glance KBM-110 Up / Down button Press briefly Long press Both simultaneously Display of the min/max value Change value of the selected parameter Reset the min/max value of the current measurement Rotate display, overhead display Function key Press briefly Freeze measurement Return to measurement display Call up next parameter Long press, 2s Start menu configuration, (ONF appears in the display Long press, 4s Start automatic calibration, (AL appears in the display 5.
KBM-110 Bases for measurement | 6 6 Bases for measurement 6.1 pH measurement 6.1.1 Explanation The pH value describes the acidic or alkaline behaviour of an aqueous solution. A pH value below 7 is acidic, a value above 7 is alkaline. A pH value of 7 is neutral. The pH measurement is very precise, but also sensitive. The measured signals are very weak and high-ohmic. This is the case, in particularly in low-ion media.
6 | Bases for measurement KBM-110 6.1.4 Further information A pH electrode is a wear part. If the signal is very slow or the required values are no longer observed after careful cleaning and possible regeneration, the electrode must be replaced. When using the electrodes, be aware that various substances in aqueous solutions can corrode glass and that chemicals can produce a chemical reaction with the KCl solution in the electrode, which can result in blockage of the diaphragm.
KBM-110 Bases for measurement | 6 6.1.7 Care and maintenance NOTE The GAK 1400 working and calibration set includes all necessary products for calibration, care and maintenance of the electrode. Normal cleaning takes place with the GRL 100 pepsin cleaning solution into which the electrode is immersed for 5 minutes before being rinsed off with clean water. NOTE Crystallisation of the 3 mol/l KCL solution is unavoidable.
7 | Maintenance KBM-110 7 Maintenance 7.1 Operating and maintenance notices NOTE The product and electrode must be handled with care and used in accordance with the technical data. Do not throw or strike. NOTE Plugs and sockets must be protected from soiling. NOTE If the product is stored at a temperature above 50 °C, or is not used for an extended period of time, the batteries must be removed. Leaks from the batteries are avoided as a result.
KBM-110 Maintenance | 7 – Only use high-quality and suitable alkaline batteries! CAUTION Damage! If the batteries have different charge levels, leaks and thus damage to the product can occur. – Use new, high-quality batteries! – Do not use different types of batteries! – Remove depleted batteries and dispose of them at a suitable collection point! NOTE Unnecessary screwing places the water-tightness of the product, among other things, at risk and should be avoided.
7 | Maintenance KBM-110 For automatic calibration, open the Calibration menu. See Automatic pH calibration [} p. 19]. Prerequisite – The pH electrode and, if applicable, a temperature sensor are inserted in the product. – The product is switched on. Instruction 1. Carefully remove the protective cap from the electrode. 2. Rinse off the electrode with distilled or deionised water. Outcome of an action Now, the product can be calibrated. 7.3.1.
KBM-110 Maintenance | 7 Ready-to-use buffer solutions in 250 ml dosing bottles with a dosing volume of 20 to 25 ml. Prerequisite – Plastic bottle – approx. 100 ml of distilled water – Buffer capsule Instruction GPH 4.0 GPH 7.0 GPH 10.0 GPH 12.0 Colour Orange Green Blue White 10 °C 3.99 7.06 10.18 12.35 20 °C 3.99 7.01 10.06 12.14 25 °C 4.01 7.00 10.01 12.00 30 °C 4.01 6.99 9.97 11.89 40 °C 4.03 6.98 9.89 11.71 Buffer capsules for 100 ml buffer solution 1. Fill a plastic bottle with approx.
7 | Maintenance KBM-110 9. Place the electrode in the second buffer solution. The product recognises whether it is a PK 4 or PK 10 buffer solution automatically. 10. If the temperature sensor is not inserted, enter the temperature of the buffer solution by pressing the Up key and Down key and confirm the entry by pressing the Function key again. 11. Then, rinse the electrode again with distilled or deionised water.
KBM-110 Prerequisite Maintenance | 7 – A buffer solution with a value between pH 6.75 and pH 7.25 is available. – A second buffer solution with a value below pH 6 and above pH 8 is available. Instruction 1. Press the Function key for 2 seconds to open the Configuration menu. 2. (ONF appears in the display. Release the Function key. 3. The parameter SET.T appears if the temperature sensor is not plugged in. If the temperature sensor is plugged in, you jump to the next point. 4.
8 | Operation KBM-110 8 Operation 8.1 Commissioning 8.1.1 Explanation The product is switched on with the On/Off button. It may be necessary to configure the product after switching on. See Configuration [} p. 22]. Description Prerequisite – Sufficiently full batteries are inserted in the product. – A suitable pH electrode is plugged in. – Press On/Off button. Instruction Outcome of an action Information about the configuration of the product appears in the display. POFF T.OF T.
KBM-110 Operation | 8 3. By briefly pressing the Function key, you can scroll through the parameters. Select the parameter you would like to configure. 4. When you have selected the desired parameter, change the parameter to the desired value with the Up button and the Down button. 5. The changes are saved after running through the entire Configuration menu. STOR appears in the display.
8 | Operation KBM-110 Setting the gradient PK.SL Current measure- Setting of the gradient for calibration of the pH ment measurement.
KBM-110 Operation | 8 Backlighting L,TE OFF Backlighting deactivated 15 30 60 120 240 Automatic shut-off of the backlighting after a selected time in seconds, during which no buttons have been pressed No automatic shut-off of the backlighting ON Factory settings IN,T Outcome of an action NO Use current configuration YES Reset product to factory settings. IN,T DONE appears in the display The changed value is saved and the Configuration menu is closed. STOR appears in the display.
8 | Operation KBM-110 Outcome of an action The Configuration menu is closed after the last parameter. NOTE If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product. 8.2.5 Configuring parameters of the adjustment menu Description The following representation shows the available parameters and various configuration options. Prerequisites The Adjustment menu is open. See Adjustment of the measuring input [} p. 25].
KBM-110 Operation | 8 NOTE If the product is switched off without saving the configuration, the last save value is reproduced on the next start-up of the product. B-H86.0.12.DB214-1.
9 | Error and system messages KBM-110 9 Error and system messages Display Meaning Possible causes Error during the last Faulty calibration calibration Battery depleted Battery depleted Remedy Conduct a new calibration BAT LO Battery depleted Replace battery (AL ERR.
KBM-110 Error and system messages | 9 Replace batteries Send in for repair B-H86.0.12.DB214-1.
10 | Disposal KBM-110 10 Disposal NOTE The device must not be disposed of with household waste. If the product is disposed of, please take it to a municipal collection point, where it will be transported to a disposal company in accordance with requirements of hazardous goods laws. Otherwise, return it to us, freight prepaid. We will then arrange for the proper and environmentally-friendly disposal. Please dispose of empty batteries at the collection points intended for this purpose. 30 / 32 B-H86.0.12.
KBM-110 Technical data | 11 11 Technical data Measuring range pH 0.00 .. 14.00 pH Redox -1500 .. +1500 mV Temperature -5 .. 105 °C Accuracy (at nominal temperature) Temperature compensation Input resistance Nominal temperature Measuring cycle Connections pH, Redox Temperature Display ± 0.02 pH ± 1 digit 1293 .. +1707 mVH ± 0.1% FS ± 1 Digit 23 .. 221 °F ± 0.3 °C Additional functions pH calibration Housing Protection rating Dimensions L*W*H [mm] and weight -5 .. 105 °C (or 23 .. 221 °F) ca.
12 | Service KBM-110 12 Service 12.1 Manufacturer If you have any questions, please do not hesitate to contact us: Contact VOLTCRAFT Distributed by Conrad Electronic SE Klaus-Conrad-Str. 1 92240 Hirschau, Germany Tel.: +49 9604 40 87 87 Fax: +49 180 5 312110 kundenservice@conrad.de WEEE reg. no. DE 28001718 32 / 32 B-H86.0.12.DB214-1.