Instructions
8 | Operation DO-400
22 / 29 B-H86.0.21.DB214-1.0
4. When you have selected the desired parameter, change the parameter to the de-
sired value with the Up button and the Down button.
5. The changes are saved after running through the entire Configuration menu. STOR
appears in the display. The Configuration menu can be exited from any arbitrary
parameter by pressing and holding the Function key for 2 seconds. The changes
made up that point are saved.
Representation
Call up menu Next parameter Change value Save changes Discard
changes
2s Press: Single
step
Hold: Rapid
change
2s 2s
Outcome of an action
The Configuration menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
8.2.3 Configuring parameters of the configuration menu
Description
The following representation shows the available parameters and various configura-
tion options.
Prerequisite
– The Configuration menu is open. See Opening the configuration menu [}p.21].
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the Up button and
Down button.
3. The available configuration options are listed for each parameter in the following
representation.
Representation
Parameter Values Meaning
Input
INP
SAT %
Oxygen saturation in per cent
(ONC mg/l
Oxygen concentration in mg/l
(ONC ppm
Oxygen concentration in ppm
Pressure
SET.P
500 .. 4000
Environmental pressure in hPa, corresponding to
mbar
Salinity correction
SAL
0 .. 70
Salinity in the measuring medium in PSU, corres-
ponding to g/kg