User's Manual

15
Recommended separation distances between portable and mobile RF communication
equipment and the Wireless ECG Monito
r
The Wireless ECG Monitor is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Wireless ECG Monitor can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Wireless ECG Monitor as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
80 MHz to 800 MHz
d = [1.17]P
800 MHz to 2.5GHz
d = [2.33]P
0.01 0.12 0.23
0.1 0.37 0.75
1 1.17 2.33
10 3.70 7.36
100 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Recommended separation distances between portable and mobile RF communications equipment and the
equipment and system – for equipment and systems that are not life supporting
8.1 EMC Compliance
This product is compliant with the electromagnetic compatibility requirements of IEC60601-1-
2.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Vpatch System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Vpatch System should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 GROUP 1 The Vpatch System uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The Vpatch System is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies