User's Manual

6
is not heavy coverage. It is important to ensure that any chest hair present does not
prevent the Biosensor Array from adhering well to the skin.
To ensure the collection of diagnostic quality ECG recordings and to reduce the
collection of “noise events” the skin MUST be cleaned using a non-alcoholic skin wipe
ensuring that the skin surface is thoroughly dried BEFORE applying the Biosensor
Array.
The Biosensor Array must be applied within two hours of skin preparation.
6.5 Biosensor Array Application
It is essential for the collection of clear event and episode recordings that a suitably
trained healthcare professional apply the Biosensor Array to the patient for the first
time.
It is the responsibility of this healthcare professional to provide education and
instruction to the patient for replacing the Biosensor Array during the monitoring
period.
The Vpod device should be placed onto the Biosensor Array after the array is applied
to the patient’s chest.
It is advisable to determine the optimum electrode placement on the patient before
removing the paper liners from the electrodes.
NOTE: The Biosensor Array is a single-use product, which is recommended for use
on the patient for a duration of up to 7 days, after which a new Biosensor Array is
to be applied to the patient’s chest.
The Biosensor Array must be applied to the patient in the configuration as illustrated
in Figure 2.
Figure 2: Biosensor Array Placement
After removing the release liners to expose the adhesive foam, it is recommended the
Health care professional instruct the patient to inhale and hold a deep breath while
they position the Biosensor Array in place. This is done to maximize patient comfort
during wear.
The health care professional must ensure that the Biosensor Array is securely fixed to
the patient’s body by smoothing each adhesive area firmly to the skin ensuring that
there are no creases.
Loose fitting outer clothing and minimal contact between the Biosensor Array and
undergarments is highly recommended.
When removing the Vpod to replace batteries or prior to bathing or showering,
remove one stud at a time while pressing down firmly on the Biosensor Array beside
each stud.