Installation guide
Installation guide 11 (of 14)
□ After installing the Dialogue:
 Switch on devices
 Run Dialogue software from Start menu (in 'Antec software')
 AFTER running Dialogue (which installs dongle drivers), insert 
the license dongle and restart the software
ROXY Potentiostat should be automatically detected (If no detector is pre-
sent, Dialogue software will continue running in 'demo' mode.)
□ Go to Options  Settings  RS232 to connect the syringe pump 
(Fig.7).
The syringe pump must be connected by user to the proper port in the 
Settings window of the Dialogue program. If the port number is not listed, 
the number can be entered manually.
Note the COM numbers that are assigned to the ROXY potentiostat 
and syringe pump can differ because the numbers are generated au-
tomatically by Windows during installation. However, it is advised to 
assign the ROXY Potentiostat with COM number 7 and syringe pump 
with COM number 8.
The instruction how to change the COM number is described in the 
next section of this document.
Fig. 7. Dialogue settings window (Options
Settings
RS232).