User manual

Chapter 6 System Configuration
VXI-MXI-2 User Manual 6-42 © National Instruments Corporation
4. Enables the logical address window of the VXI-MXI-2 found at
logical address 82 for the entire outward mapping range of 0 to FF.
Scans all logical addresses, skipping all previously encountered
devices, and finds the VXI-MXI-2 in VXIbus Mainframe #4 and
the VXI-MXI-2 in VXIbus Mainframe #5.
5. Enables the logical address window of the VXI-MXI-2 in VXIbus
Mainframe #4 for the entire inward mapping range of 0 to FF.
Scans all logical addresses, skipping all previously encountered
devices. Finds the Slot 0 device and uses it to move all DC devices
in VXIbus Mainframe #4 to the lowest unused logical addresses.
No more VXI-MXI-2 interfaces are found. The RM enables the
logical address window for the VXI-MXI-2 in VXIbus
Mainframe #4 with an inward range of A0 to AF hex by writing
the value 64A0 hex to the Logical Address Window Register
(Base/Size format).
6. Enables the logical address window of the VXI-MXI-2 in VXIbus
Mainframe #5 for the entire inward mapping range of 0 to FF.
Scans all logical addresses, skipping all previously encountered
devices and previously defined address ranges. Finds the Slot 0
device and uses it to move all DC devices in VXIbus
Mainframe #5 to the lowest unused logical addresses. No more
VXI-MXI-2 interfaces are found. The RM enables the logical
address window for the VXI-MXI-2 in VXIbus Mainframe #5 with
an inward range of B0 to BF hex by writing the value 64B0 hex to
the Logical Address Window Register (Base/Size format).
7. Sets the logical address window of the VXI-MXI-2 found at
logical address 82 to cover the ranges of the VXI-MXI-2 in
VXIbus Mainframe #4 (A0 to AF) and the VXI-MXI-2 in VXIbus
Mainframe #5 (B0 to BF). Enables the logical address window of
the VXI-MXI-2 at logical address 82 with an outward range of A0
to BF by writing the value 43A0 hex to the Logical Address
Window Register (Base/Size format).
8. Sets the logical address window of the VXI-MXI-2 found in
VXIbus Mainframe #3 at logical address 80 to cover the devices in
that mainframe (80 to 8F) and the ranges required of its Level 2
devices: the VXI-MXI-2 in VXIbus Mainframe #4 (A0 to AF) and
the VXI-MXI-2 in VXIbus Mainframe #5 (B0 to BF). Enables the
logical address window of the VXI-MXI-2 at logical address 80
with an inward range of 80 to BF by writing the value 6280 hex to
the Logical Address Window Register (Base/Size format).