User manual

Chapter 7 VXIplug&play for the VXI-MXI-2
© National Instruments Corporation 7-13 VXI-MXI-2 User Manual
CLK10
The VXI-MXI-2 can either receive or drive the MXIbus CLK10 signal.
In the default setting of Switch, the C-size VXI-MXI-2 uses the switch
setting of S7 for this determination; use switch S1 if you have a
VXI-MXI-2/B.
You can use the Drive or Receive options of the CLK10 feature to
override the switch setting and control the direction of the MXIbus
CLK10 signal. When receiving the MXIbus CLK10 signal, configure
the W3 jumper setting on the C-size VXI-MXI-2 (or the W1 jumper
setting on the VXI-MXI-2/B) to use the MXIbus as the source for
generating the VXIbus CLK10 (applicable only if the VXI-MXI-2 is a
Slot 0 device). When driving the MXIbus CLK10, the VXIbus CLK10
is used as the source. In this case, change the jumper setting so that it
does not use the MXIbus CLK10 as the source for the VXIbus CLK10.
Warning: Do not configure more than one MXIbus device to drive the MXIbus
CLK10. Setting up a second device to drive MXIbus CLK10 could
damage the device.
VXI-MXI-2 VXI
plug&play
Knowledge Base File
A VXIplug&play Knowledge Base File is included on the
VXIplug&play disk in this kit. This file conforms to VPP-5, VXI
Component Knowledge Base Specification. This file contains detailed
information about the VXI-MXI-2 such as address space requirements
and power consumption. The knowledge base file is intended to be
used with software tools that aid in system design, integration, and
verification.
The knowledge base file is directly accessible on the disk as a text file
with a .kb extension. It is also installed on your system when the
Setup program is executed as described earlier in this chapter. The
knowledge base file is directly accessible for users who are not running
Windows and cannot use the Setup program to install the file.