Appendix E Configuring a Two-Frame System
© National Instruments Corporation E-7 VXI-MXI-2 User Manual
To configure the Frame A VXI-MXI-2 to drive the MXIbus CLK10,
change the setting of S7 as shown in Figure E-6b if you have a C-size
VXI-MXI-2, or change the setting of S1 as shown in Figure E-7b if
you have a VXI-MXI-2/B. The setting of the W3 jumper (W1 on a
VXI-MXI-2/B) does not matter because the VXI-MXI-2 is not in Slot 0
and will not be driving the VXIbus CLK10.
To configure the Frame B VXI-MXI-2 to route the MXIbus CLK10 to
the VXIbus, keep S7 in its default setting as shown in Figure E-6a, or
see Figure 7a for the default setting of S1 if you have a VXI-MXI-2/B.
Change the setting of the W3 jumper on the C-size VXI-MXI-2 as
shown in Figure E-8. If you have a VXI-MXI-2/B, see Figure E-9 for
the proper setting of the W1 jumper.
a. Receive CLK10 from MXIbus (Default) b. Drive MXIbus CLK10 from VXIbus CLK10
S7
Drive CLK10 out MXIbus
Receive CLK10 from MXIbus
S7
Drive CLK10 out MXIbus
Receive CLK10 from MXIbus
Figure E-6. Receiving or Driving MXIbus CLK10 on a C-Size VXI-MXI-2
a. Receive CLK10 from MXIbus (Default) b. Drive MXIbus CLK10 from VXIbus CLK10
S1MBCLK10
IN OUT
S1MBCLK10
IN OUT
Figure E-7. Receiving or Driving MXIbus CLK10 on a VXI-MXI-2/B
Warning: Do not configure more than one MXIbus device to drive the MXIbus
CLK10. Having a second device driving MXIbus CLK10 could damage
the device.