User`s manual

www.vxitech.com
Tektronix to VXI Technology Cross-Ref. Guide 109
SECTION 12
VX4101A TO VTI4101A: DMM/COUNTER
SPECIFICATION COMPARISONS (DMM)
Tektronix VX4101A: 5.5 Digit DMM VXI Technology Inc. VTI4101A: 6.5 Digit DMM
DIGITS APERTURE READINGS/SEC DIGITS APERTURE READINGS/SEC
6.5 Not Available N/A
6.5
166.7 ms to 2 s 2
5.5 16.67 ms < 60 5.5 1.667 ms to 20 ms 150
4.5 < 16.67 ms > 60 4.5 0.5 ms 500
DESCRIPTION DESCRIPTION
4096 Measurements 64,000 measurements (256,000 Optional)
RANGE MAX READING RESOLUTION RANGE MAX READING RESOLUTION
30 mV ± 30.0000 mV 100 nV 100 mV ± 120 mV 100 nV
300 mV ± 300.000 mV 1 µV 1 V ± 1.2 V 1 µV
3 V ± 3.00000 V 10 µV 10 V ± 12 V 10 µV
30 V ±-30.0000 V 100 µV 100 V ± 120 V 100 µV
300 V ±300.000 V 1 mV 300 V ± 300 V 100 µV
RANGE RESISTANCE RANGE RESISTANCE
30 mV, 300 mV, 3 V 10 M ± 5% or >10 G 100 mV, 1 V and 10 V 1 G
30 V, 300 V 10 M ±5% 100 V and 300 V 10 M
RANGE MAX READING RESOLUTION RANGE MAX READING RESOLUTION
30 mV 30.0000 mV 100 nV 100 mV ac 120 mV ac 1 µV
300 mV 300.000 mV 1 µV 1 V ac 1.2 V ac 10 µV
3 V 3.00000 V 10 µV 10 V ac 12 V ac 100 µV
30 V 30.0000 V 100 µV 100 V ac 120 V ac 1 mV
300 V 300.000 V 1 mV 300 V ac 300 V ac 1 mV
RANGE MAX READING RESOLUTION RANGE MAX READING RESOLUTION
30 30 100 µ 20 20 100 µ
300 300 1 m 200 200 100 µ
3 k 3 k 10 m 2 k 2 k 1 m
30 k 30 k 100 m 20 k 20 k 10 m
300 k 300 k 1 200 k 200 k 100 m
3 M 3 M 10 2 M 2 M 1
30 M 30 M 100 20 M 20 M 100
300 M 300 M 1 k (nominal)
RANGE MAX READING RESOLUTION RANGE MAX READING RESOLUTION
150 mA ±150.0000 mA 0.5 µA 10 mA 12 mA 10 nA
1.0 A ±1.00000 A 5 µA 100 mA 120 mA 100 nA
1.0 A 1.2 A 1 µA