Universal Remote User Manual

VT1422A Command Reference 313Chapter 7
OUTPut
Parameters
Comments Select 488E-6 (or MAX) for measuring resistances of less than 8000 Ω. Select
30E-6 (or MIN) for resistances of 8000 Ω and above. <amplitude> may be
specified in µA (ua).
For resistance temperature measurements ([SENSe:]FUNCtion:TEMPerature)
the Current Source SCP must be set as follows:
When *CAL? is executed, the current sources are calibrated on the range
selected at that time.
When Accepted: Not while INITiated
Related Commands: *CAL?, OUTP:CURR:AMPL?
*RST Condition: MIN
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage OUTP:CURR:AMPL 488ua,(@116:123)
Set Current Source SCP at channels 16
through 23 to 488 µA
OUTP:CURR:AMPL 30E-6,(@105)
Set Current Source SCP at channel 5 to
30 µA
OUTPut:CURRent:AMPLitude?
OUTPut:CURRent:AMPLitude? (@<channel>) returns the range setting of the
Current Source SCP channel specified by channel.
Parameters
Comments The <channel> parameter must specify a single channel only.
If <channel> specifies an SCP which is not a Current Source, a +3007, "Invalid
signal conditioning plug-on" error is generated.
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
amplitude numeric (float32) MIN | 30E-6 | MAX | 488E-6 A dc
ch_list channel list (string) 100 - 163 none
MAX (488 μA)
for RTD,85 | 92 and THER,2250
MIN (30 μA)
for THER,5000 | 10000
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel channel list (string) 100 - 163 none