Parameter Settings
Many parameters are channel-dependent, meaning that each channel can be
set independently of the others in the module. Other parameters are
module-dependent; changing a module-dependent parameter for a channel
will change it for all channels on that module. For example, changing
blocksize, a module-dependent parameter, for input channel 3 will also
change the block size for all other channels in the same Agilent E1432A
module as channel 3.
When possible, parameters are written to the hardware as soon as they are
received. Sometimes, the parameter can’t be written to the hardware until
the start of a measurement; in this case the value of the parameter is saved
in RAM in the Agilent E1432A module until the measurement is started
with e1432_init_measure. Some parameters can be changed while a
measurement is running, but many do not take effect until the next start of
a measurement.
Measurement Initiation
This section describes the measurement initiation process in the Agilent
E1432A.
The measurement initialization states, and the corresponding Sync/Trigger
line transitions (with ‘H’ for high, ‘L’ for Low) are:
The module enters the TESTED state after a reset. In this state, all of the
module parameters may be set. The Agilent E1432A stays in the TESTED
state until it sees a high-to-low transition of the Sync/Trigger line.
LHLHL HLH
Sync/Trigger line
Tes ted
Booting Pre-armBooted
Settling
Idle
Figure 4-4: Measurement initialization
Agilent E1432A User's Guide
The C-Language Host Interface Library
4-20