Owner`s manual
Calibration
OPERATION
Enter the Menu Selection function and
activate the Calibration item.
DISPLAY
Check Calibratn
xx.x %MC
STORE
Press the any button to return to the
%MC Readings mode.
GROUP
MENU
ARROWS
Checking Calibration
Calibration is factory set. It may be checked by using a
Wagner Calibration Verification Block (CVB) (P/N: 840-
60130-002). Use only to verify L610 factory calibration is
within tolerance limits. If calibration is out of tolerance,
return L610 to factory or repair depot for re-calibration.
Never adjust L610 to this device. Protect CVB from
direct sunlight. Store in cool dry place. Follow these
steps for checking calibration.
1. Enter the Calibration menu item.
2. Hold the meter in mid air so the bottom is several feet
from any object. The correct reading for air is .0
%MC. If the reading is outside the range, -2 to 2
%MC, the unit is out of calibration.
3. Place the CVB, rubber feet down, on a metal surface:
such as a file cabinet, a piece of sheet metal, etc.
4. Orientate the long axis of the handmeter with the long
axis of the CVB. Press the handmeter sensor in the
center of the CVB and observe the reading. The
correct reading is 17.5 %MC. If the reading is outside
the range, 15.5 to 19.5 %MC, the unit is out of
calibration.
Warning—There are two (2) hole plugs located on either
side of the L610 handle. Do not remove these plugs
because the adjustments below them are factory set and
should not be adjusted by the customer. If adjustment is
attempted, the unit will not be in calibration and will
require factory re-calibration. Also if adjustment is
attempted, warranty will be voided.
500-61001-001 Rev. F Page 15 of 30