Product Manual

For All Star Colposcope Models
CLASSIFICATION
Model Safety Class
Star with Trulight Colposcopes
I
ALL models of Colposcopes have no Applied Parts.
Do not get fluid into the Colposcope. Should any liquid or solid object fall into the unit,
unplug the unit and call Technical Support (866) 928-3211.
The LED Model Colposcopes are suitable for continuous operation.
The LED Model Colposcopes are classified as normal equipment (IPX0) according to
protection against ingress of water.
ENVIRONMENT CONDITIONS
Use
Environmental Temperature:
Relative Humidity:
Air Pressure:
between +10 °C and +40 °C
between 10% and 90%
between 700 hPa and 1060 hPa
Shipping and Storage:
Environmental Temperature:
Relative Humidity:
Air Pressure:
between +10 °C and +40 °C
between 10% and 90%
between 700 hPa and 1060 hPa
LED LIGHT SPECIFICATION
White LED:
Correlated Color Temperature 3000-3200K
Rendering Index 90 min
Radiometric Power 80 mW
Total Irradiance Less than 1 W/m
at max intensity
Lifetime [max. drive I=1000mA] Min. 25,000 hours
Green LED:
Average Wavelength 520-535 nM
Radiometric Power 80 mW
Total Irradiance
Less than 1 W/m
2
at max intensity
IMSC036-DFU • Rev. B • 10/14 15 Made in the USA