Installation Manual

HF3000-Series Installation Manual
Doc. No. Revision
07-0002 Rev.3
Page 2 of 14
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
1 PURPOSE
This procedure describes the specifications, installation steps, and regulatory information for
the HF3000-series SmartCabinet.
2 SCOPE
This procedure applies to installation of HF3000-series SmartCabinets
3 REFERENCE DOCUMENTS
None
4 PRODUCT SPECIFICATION:
Power Brick
100-240 VAC, 1.5A Max, 50-60 Hz, power supply is rated to IEC/EN/UL-62368-1
and IEC/EN/UL-60601-1 (medical grade).
Cabinet Power
AC Cable
12.5V DC, 7.2A
The cabinet is supplied with a power cable.
Fuse
None
Environment
Operating environment between 50 and 90 degrees F (10C 32C), non-
condensing; must be installed in an environment free of corrosive agents and
excessive dust
Size
78 inches high by 36 inches wide by 27 inches deep
Weight
Weight in lbs.
Without Doors
Base Cabinet with 2 shelves
160
Cabinet with 3 shelves
175
Cabinet with 4 shelves
190
Cabinet with 5 shelves
205
Cabinet with 6 shelves
220
Cabinet with 7 shelves
235
Cabinet with 8 shelves
250
Cabinet with 9 shelves
265
Cabinet with 10 shelves
280
Cabinet with 11 shelves
295
Weight Capacity
The weight capacity of each shelf if 25 lbs.
Network
Auto-select 10/100Mbps Ethernet