-200,"Execution error". This is the generic syntax error for the
instrument when it cannot detect more specific errors. Execution error
as defined in IEEE-488.2 has occurred.
-221,"Setting conflict". Two conflicting parameters were received
which cannot be executed without generating an error. Listed below
are events causing setting conflicts.
1. Sum of pulse or ramp parameters is more than 100. Corrective
action: Change parameters to correct the problem.
2. ampl/2 + |offset| is more than 2.5. Corrective action: Reduce offset
to 0, then change amplitude-offset values to correct the problem.
3. Activating filters when the 302 is set to output the built-in sine
waveform, or activating the built-in sine waveform when one of the
302 filters is turned on. Corrective action: If in sine, select another
function and activate the filter(s).
4. Activating burst mode when the 302 is set to sequence mode, or
activating sequence mode when the 302 is set to burst mode.
Corrective action: Remove the 302 from burst or sequence and
then selected the desired mode.
5. An attempt to change sequence mode while the 302 is set to
operate in one of the sequence modes. Corrective action: remove
the 302 from sequence mode, select the required sequence type,
then resume sequence operation.
6. Changing operating mode from triggered to continuous when the
302 is set to single sequence advance, or changing the operating
mode from continuous to triggered when the 302 is set to
automatic sequence advance mode. Corrective action: Observe the
302 advance mode while setting sequence advance.
7. In multi-sequence operation, trying to select a sequence number
larger than the specified number of sequences. Corrective action: If
larger number of sequences is required, exit sequence mode,
select a larger number of sequences and then resume sequence
operation.
8. An attempt to change the sequence number but the 302 is in single
sequence mode. Corrective action: Exit sequence mode, select the
multi-sequence mode, select the number of sequences you are
about to use and then resume sequence operation.
-222,”Data out of range”. Parameter data, which followed a specific
header, could not be used because its value is outside the valid range