Figure 3.9 Adjusting Parameters of standard signals
Changing the
Frequency and
Sample Clock
The user should be careful not to confuse waveform frequency with
sample clock frequency. The waveform frequency parameter is valid
for standard waveforms only and controls waveform frequency at the
output connector. On the other hand, the sample clock frequency
parameter is valid for arbitrary waveforms only and defines the
frequency of which the generator clocks data points.
Standard waveform frequency is measured in units of Hz. Arbitrary
waveform sample clock frequency is measured in units of Sa/s
(samples per second). The frequency of a given arbitrary waveform at
the output connector is dependant on sample clock frequency, the
number of data points, and the specifics of the waveform defined
Use the following command to change the frequency for standard
waveforms:
FREQuency {<frequency> | MINimum | MAXimum}
MIN selects the lowest frequency allowed for the currently active
function. MAX selects the highest frequency allowed for the currently
active function. The default frequency setting is 1 MHz for all
parameters.
The query:
Configuring the Instrument
31