Specifications

User Manual 301/302
FREQuency?
queries the frequency setting for the standard function currently active
and returns a value in hertz.
The command:
FREQuency:RASTer {<frequency> | MINimum |MAXimum}
Sets the sample clock frequency for the user and sequenced mode
parameters. MIN selects the lowest frequency allowed for the currently
active segment or sequence. MAX selects the highest frequency
allowed for the currently active segment or sequence. The default
sample clock frequency setting is 10 MHz for all.
The query:
FREQuency:RASTer?
queries the sample clock frequency setting for the arbitrary segment
or sequence currently active and returns a value in Hertz.
Using WaveWiz, frequency of standard signals is also set in the
Standard tab in the Instrument Control Panel. See Figure 3.9 above.
Selecting the
External SCLK
In cases where synchronization to other instrument in the system is
needed, one may use the front panel SCLK input for replacing the
internal sample clock generators. Use the following command to
select the external SCLK input:
FREQuency:RASTer:SOURce {INTernal | EXTernal}
INT is the default state of the instrument after power up. Select EXT to
activate the front panel SCLK input.
The query:
FREQuency:RASTer:SOURce?
queries the sample clock source and returns INT or EXT.
WaveWiz has no specific command control for this function; it is set by
sending a SCPI command directly from the system menu in the
Instrument Front Panel. (For a more detailed description see later in
this chapter).
Configuring The Instrument