Specifications

User Manual 301/302
and select Sequenced output.
NOTE
The sequence generator goes through its steps in ascending order. In
continuous mode, the sequence is repeated automatically after
completing the last step. In Triggered mode, the output stops at the
last point of the last waveform. In Gated mode, the sequence is
always completed after the gate stop signal.
Adjusting Phase
Offset Between
Channels
The prime purpose of channel synchronization is to control the start
phase in such a way that all channels start generating waveforms at
the same instance, regardless of their respective waveform shape and
frequency.
In general the initial start phase offset is around 0°. However, due to
variations in circuit design and propagation delays, there is an
unavoidable (but minimal) skew between the channels. The initial
skew between channels is specified in Appendix A.
There are no internal provisions to correct skew between channels
however, it can be easily corrected by adjusting the length of the
cables that are connected to the output connectors. One meter of 50
coax cable amounts to about 2 ns of delay time. Use this general
guideline to adjust the skew for your system. For example, if channel 2
lags 1.2 ns after channel 3, you have to delay the output of channel 3
by 1.2 ns. You’ll have to extend the cable for channel 3 by roughly 0.6
meter to compensate for this skew.
Phase offsets are always referenced to channel A. Therefore, you can
program phase offsets for channels B only. An attempt to program
channel A with this parameter will result in an error message.
There are two ways to program phase offsets: 1) Using an angle and
2) Using waveform points. If you use angles to program your phase
offset, you should first compute the equation given in Appendix A to
determine your phase resolution and error. Using waveform points to
specify phase offset is much easier because you have full control over
waveform length and can easily determine the phase offset from the
number of waveform points offset.
TIP
Phase offset between channels is programmable with resolution o
f
1 waveform point. The more points you have for your waveform, the
better the phase offset resolution you can have. For example, for a
1,000-point waveform, your resolution is 1 / 1000 x 360 = 0.36°; for a
Configuring the Instrument
73