User manual

Detailed descriptions of I/O modules | Analogue input module UR20-4AI-TC-DIAG
209Manual u-remote1432790000/03/02.2014
Overview of the editable parameters UR20-4AI-TC-DIAG
Channel Description Options(Value) Default value
Temperature unit Degree Celsius (0) / Degree Fahrenheit (1) / Kelvin (2) Degree Celsius
0 ... 3 Measurement range TC Type J (0) / TC Type K (1) / TC Type N (2) / TC Type R (3) / TC Type S (4) /
TC Type T (5) / TC Type B (6) / TC Type C (7) / TC Type E (8) / TC Type L (9) /
TC Type U (10) /
± 15,625 mV (11) / ± 31,25 mV (12) / ± 62,5 mV (13) / ± 125 mV (14) /
± 250 mV (15) / ± 500 mV (16) / ± 1000 mV (17) / ± 2000 mV (18) /
disabled (19)
disabled
0 ... 3 Cold junction compensation internal (0) / external Channel 0 (1) / external Channel 1 (2) /
external Channel 2 (3) / external Channel 3 (4)
internal
0 ... 3 Conversion time 240 ms (0) / 130 ms (1) / 80 ms (2) / 55 ms (3) / 43 ms (4) / 36 ms (5) 80 ms
0 ... 3 Channel diagnostics disabled (0) / enabled (1) disabled
0 ... 3 Limit value monitoring disabled (0) / enabled (1) disabled
0 ... 3 High limit value -32768 ... 32767 0
0 ... 3 Low limit value -32768 ... 32767 0