[F10] Display mode
Select instantaneous flow or accumulated flow to be displayed.
<Operation>
Press the button.
Press the button to set.
[F10] Setting of display mode completed.
Return to function selection mode.
The accumulation flow count will start when power is supplied.
Accumulated flow can be displayed up to 999999999 L (9999999999 ft
3
).
The position of the dots indicates which digits are displayed.
The first 3 digits of the measurement value are always displayed (It changes automatically).
Press the and buttons to display the other digits.
Accumulated value
Upper 3 digits
Middle 3 digits
Lower 3 digits
50 L
button
button
Always displayed
60,010 L
button
Always displayed
button
203,481,020 L
Always displayed
button
button
The accumulated value can be reset as follows:
Accumulated flow reset (page 54).
Accumulated flow external reset (page 44).
Turn the power ON (this will not reset the value if accumulated value hold is selected, refer to page 46).
Display mode
[FLo] and [the current set value] are displayed in turn.
Press the or button to select.
Press the or button in function selection mode to display [F10].
[F10] and [FLo] are displayed in turn.