Owner's manual

27
i=2
On Off Off
Õ Ù â á
Adjust Output Current to 2 mA
i=2 Ó
Accept output
i=20 Õ Ù â á
Adjust Output Current to 20 mA
i=20 Ó
Accept output
Sure Ó
To Save Values (or any other key to restart).
Current Sink Mode
Display I+ I- V Keys Action
Switch on Power, Calibrate link In CAL Position
u=0
Off Off On
Ò Ò
Press twice to select Current Sink mode
Connect Power Supply and Multimeter to measure output current (mA).
i=2
Off On Off
Õ Ù â á
Adjust Output Current to 2 mA
i=2 Ó
Accept output
i=20 Õ Ù â á
Adjust Output Current to 20 mA
i=20 Ó
Accept output
Sure Ó
To Save Values (or any other key to restart).
External Voltage Inputs
Display I+ I- V Keys Action
Note: you must calibrate the Voltage Source Mode before completing this procedure.
Switch on Power, Calibrate link In CAL Position
u=0
Off Off On
Ô Õ Ù Ù Ó
Press sequence to select External Voltage
mode. INC and DEC Leds will both light.
Connect voltage source and Set to 0 V.
U=0
Off Off On
Ó
Calibrate 0 V point
Current Source Mode (continued)