Manual

WI-I/O 9-x_WI-I/O-EX-1-S-x 2.16 Page 31
In general, the following may be used as a rule of thumb for calculating the appropriate sensitivity
required for a given application:
Instantaneous change of 2 x sensitivity on input 3 second output response
Instantaneous change of 10 x sensitivity on input 5 second output response
The analog inputs have 15 bit resolution and 0.016mA accuracy.
Pulse input change of state
Pulse input counts do not use “change-of-state” transmissions. Instead, accumulated pulse input
counts are transmitted at set intervals. The default period is 1 minute and is configurable. The absolute
pulse count is transmitted. If the PI is transmitted to a PO on a WI-I/O 9-x_WI-I/O-EX-1-S-x module,
then the pulse outputs are re-created from the accumulated pulse count. Rollovers of the pulse count
thru zero are catered for. If a transmission is missed, the pulse output will still be re-created when the
next accumulated value is transmitted. This ensures that no pulses are lost due to communications
failures. If the PI is transmitted to a WI-I/O 9-C interface module, then the accumulated pulse count is
stored in the WI-I/O 9-C for interfacing to the host device.
The following diagram shows how pulse inputs are re-created as pulse outputs. For pulse outputs, the
module keeps two counters in memory - the pulse input count received from the remote module, and
the count of output pulses. When the module receives an update of the input pulse count, it will output
pulses until the output pulse count is the same as the input pulse count. The output pulse will be output
evenly over the pulse output update time which is configured in the module. For example, assume that
module receives a pulse input update message from the remote module, and the difference between the
pulse input count and the pulse output count is 12 pulses. The module will then output the 12 pulses
evenly over the next minute (if the pulse output update time is 1 minute).
Input
Pulses
Time
Output
Pulses
PI update time
PO update time
Time
The default values for the pulse input update time and pulse output update time is 1 minute. In this
case, the output pulses are effectively 1 minute behind the input pulses. These update times may be