The following features and selections are available in the Biotech mode.
Table 2- 3: Biotech Mode Features
Function Setting
DEVICE VERSION The display will show “X.X”, where “X.X” is the device software version number (i.e.
1.0). This is helpful when discussing operation with Welch Allyn Customer Care.
ALGORITHM
VERSION
The display should show “X.X” where “X.X” is the mode Software Version (i.e. 7.5).
This is helpful when discussing operation with Welch Allyn Customer Care.
ID
(Model 692 Only)
This section of the display is designed to store up to a seven digit alpha-numeric
thermometer ID. To store the thermometer's ID, press and release either the Mode
or the C/F button to change the value of the flashing digit. Press the Pulse Timer
button to move to the next digit.
DEFAULT MODE You may choose between four settings as the default predictive mode: Oral, Adult
Axillary, Pediatric Axillary, and Last (last prediction). To change the default mode,
press and release the C/F button or the Mode Selection button to advance to the
next mode. This setting is retained in non-volatile memory when the batteries are
replaced.
UNITS DEFAULT This section displays the temperature measurement scale. To change the
measurement scale from °F to °C, press and release the C/F button or Mode
Selection button. This setting is retained in non-volatile memory when the batteries
are replaced.
SITE INDICATION Three options are available for mode indication: icon and words, words only, and
icon only. To change the setting, press and release the Mode Selection button or C/F
button to advance to the next type of indication. This setting is retained in non-
volatile memory when the batteries are replaced.
SECURITY COUNT
(Model 692 Only)
The anti-theft function uses a temperature counter that can be set at OFF, 25, 50,
100, or 200.
Press and release the Mode Selection button to select the desired security
count.
Press and release the C/F button to select descending security count option (i.e.
200, 100, 50, 25, OFF).
The instrument is set to the selected security count after the probe is returned to the
probe well. This setting is retained in non-volatile memory when the batteries are
replaced.
INSTANT ON
(Model 692 Only)
The thermometer can be set to alarm when removed from the wall holder. This
alarm can be disabled by the user by simultaneously pressing the Mode
Selection button while pulling the probe from the probe well. When this security
feature is set to ON and the thermometer is returned to the wall holder the
security feature is reactivated.
To set this security feature, press and release the Mode Selection button or C/F
button to select ON (activated) or OFF (deactivated). This setting is retained in
non-volatile memory when the batteries are replaced.
BATTERY This setting displays the current battery voltage with 10 mV resolution. The battery is
considered acceptable if it measures higher than 3.4 volts. A new battery should
produce 4.5 volts or more.
COUNTER
Predictive Temperature
Counter
This setting displays the number of predictive temperature measurements that have
been completed (in 100's).
ERROR LOG This setting displays the last 10 saved error messages in a last-in, first-viewed
sequence. This is helpful when discussing operation with Welch Allyn Customer
Care. If there are no error messages, the display shows EMPTY LOG.
CLEAR LOG This setting appears if the thermometer has logged any error messages. Press and
release the Mode Selection button to select YES to clear the error log or NO to keep
the entries in the log.