CompacVideo Otoscope Model 23120 (NTSC) and 23120P (PAL) Operating Instruction Manual
Thank you for purchasing the Welch Allyn CompacVideo Otoscope System for ear care professionals. The operating and maintenance instructions found in this manual should be followed to ensure years of reliable service. Please read these instructions thoroughly before using your CompacVideo Otoscope System. If you have questions regarding any Welch Allyn products please call Customer Service at (800)535-6663 C US N344 ETL listed: UL2601-1, CSA C22.2 No. 601.1, IEC 60601-1, AS 3200.1.
electronic or mechanical, including photocopy, without prior consent in writing from Welch Allyn. Printed in the U.S.A.
Table of Contents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Safety & Warnings. . . . . . . . . . . . . . . . . . . . . . . 2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Using The CompacVideo Otoscope System. . . 5 Lamp Replacement . . . . . . . . . . . . . . . . . . . . . . 5 Battery Replacement. . . . . . . . . . . . . . . . . . . . . 6 Specifications . . . . . . . . . . . . . .
Introduction The Welch Allyn CompacVideo Otoscope System is a portable, battery operated video camera to be used for obtaining high quality video images of the ear canal and tympanic membrane for the ear care professional. The CompacVideo Otoscope System consists of a video head (A) and an RCA Video Cable (B). A 3.5 volt Welch Allyn battery powered handle and a video monitor are necessary for the on screen display of video images.
Symbols Consult operator’s manual for important information Type BF IPXØ Not protected against the ingress of water Safety & Warnings DO NOT USE IN THE PRESENCE OF FLAMABLE ANESTHETICS In order to obtain optimum video performance and ensure patient safety, it is important that this booklet be read thoroughly prior to using the instrument. If you have any technical or clinical questions concerning the use and / or care of the Video System, please call our Customer Service Department at (800)535-6663.
This product has no user serviceable/replaceable parts except for the lamp and battery. Please refer all other service to authorized service personnel. NOTE: This product complies with current required standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in close proximity to other equipment.
4. Attach a Welch Allyn (4 mm) speculum, provided with the system, to the video head by rotating clockwise. The quality of the video image will deteriorate if other speculum are used. 5. Turn on the power switch on the monitor. 6. Turn the rheostat on the power handle to FULL POWER. For optimum performance, the Video System should always be used with the rheostat set on maximum brightness. The electronic controls within the Video System will automatically optimize the brightness of the image.
Using The CompacVideo Otoscope System The CompacVideo Otoscope System allows the visual anatomy to be shared with a patient. Indications for use include: • Inspection of the outer ear canal • Inspection of the tympanic membrane • Inspection of hearing aid devices The CompacVideo Otoscope System should not be used for any other purposes.
Battery Replacement Caution: The battery handle must be disconnected from the supply when removing batteries. Non-Convertible Handle R ECH ARGEA BL E BA TTERY 3 .5 VOLTS CATA LOG NO. 7 23 0 0 When a rechargeable battery will no longer accept a charge, replace as follows: 1. Unscrew bottom cap. 2. Shake out old battery. 3. Insert replacement battery Important: Please note battery polarity (see figure). Caution: Insert battery fully until seated.
Specifications Environment Transport/Storage: -20°C - 49°C, 10% - 95% R.H. Max, 500hPa - 1060hPa Altitude Operating: 10°C - 40°C, 30% - 75% R.H. Max, 500hPa - 1060hPa Altitude Internally Powered Equipment Duty Cycle Operation To minimize external otoscope housing temperatures, on-time should not exceed 90 seconds and off-time should not be less than 10 minutes.
Replacement Parts: Part Number 8 Description 23101 CompacVideo Otoscope System (NTSC) - Includes VideoOtoscope Head, 10 Disposable (4 mm) Specula, Video Cable, and 71000-A Power Handle 23100 CompacVideo Otoscope (NTSC) - Includes Video Otoscope Head, 10 Disposable (4 mm) Specula, and Video Cable 23101P CompacVideo Otoscope System (PAL) - Includes VideoOtoscope head, 10 Disposable (4 mm) Specula, Video Cable, and 71000-A Power Handle 23100P CompacVideo Otoscope (PAL) - Includes Video Otoscope Head,
Limited Warranty One Year Limited Warranty On New Instruments Instrumentation purchased new from Welch Allyn, Inc. (Welch Allyn) is warranted to be free from original defects in the material and workmanship under normal use and service for a period of one year from the date of first shipment from Welch Allyn. This warranty shall be fulfilled by Welch Allyn or its authorized representative repairing or replacing at Welch Allyn’s discretion, any such defect, free of charge for parts and labor.
10
Welch Allyn, Inc. 4341 State Street Road Skaneateles, NY 13153 Tel: (800) 535-6663 Printed in U.S.A. Part No. 231101 Rev.