User Manual Owner manual

Directions for Use Chapter 9 Compliance 117
Electromagnetic compatibility
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment. The Propaq CS monitor complies with IEC EN 60601-1-
2:2001.
All medical electrical equipment must be installed and put into service in
accordance with the EMC information provided in this document and in Propaq
Directions for Use.
Portable and mobile RF communications equipment can affect the behavior of
medical electrical equipment.
Propaq CS monitors comply with all applicable and required standards for electromagnetic
interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is safe to operate the monitor in the presence of high-frequency surgical
equipment.
However, it is good practice to avoid using the monitor in extremely close
proximity to other equipment.
Propaq CS monitor: Guidance and manufacturer’s declaration—Electromagnetic emissions
The Propaq CS monitor is intended for use in the electromagnetic environment specified below. The customer or the user
of the monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment—guidance
RF emissions
CISPR 11
Group 2 The Monitor must emit electromagnetic energy in order to
perform its intended function. Nearby electronic equipment
may be affected.
RF emissions
CISPR 11
Class B The monitor is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies