Directions for Use Acuity 73
If the Wireless Propaq CS is moved out of range or loses communication with the FlexNet
network and Acuity, it continues to monitor the patient and display patient data. While not
communicating with Acuity, the Propaq CS continues to generate local patient alarms or
alert messages. Acuity does not perform waveform analysis or generate arrhythmia or ST
analysis messages while the Propaq CS is not communicating with Acuity. When the
Wireless Propaq CS is returned to within range of the FlexNet network, it automatically
reconnects to Acuity and uploads trend information.
A Propaq CS with the Acuity hardwire communication option can be identified by the
Acuity connector on the right side panel.
A Wireless Propaq CS can be identified by the yellow antenna cap on the top right corner
of the monitor in front of the carrying handle. Every Wireless Propaq CS includes the
Acuity hardwire communication option.
If a Wireless Propaq CS monitor is communicating with the network via the wireless
interface and then the Acuity hardwire cable is used to connect the bedside Acuity
network jack and the monitor, the monitor disconnects from the wireless connection and
communicates through the hardwire connection. There may be a short interruption in the
display of patient waveforms at Acuity while this change occurs.
Note
The Wireless Propaq CS can not be used with the Modem-Propaq
communication option, even when the Wireless Propaq CS is being used with a
hardwire connection or is out-of-range of a FlexNet Network access point.
DEFIB SYNCHROMONITOR
EKG x 1000
!
3A
2AG
!
!
!
12-28V, 3A
Right Side Panel
Acuity hardwire connector
Wireless Propaq CS
Yellow antenna cap indicates
Wireless Propaq CS