User Manual Owner manual

8
91
Reference
Set the time and date
1. From the Main Menu press SETUP, MORE, MORE to display the Time/Day window:
2. Press NEXT, UP, and DOWN as needed to set the time and date. Then press ENTER
to store the new time and date.
Time/Day settings and trends
When you change the hour/minute/second setting for the monitor in the Time/Day
window, the monitor deletes any patient trend data that is older than five hours for non-
NIBP trends or older than eight hours for NIBP trends according to the new clock setting.
However, if the monitor has not yet stored the full capacity of trends and you change the
hour/minute/second setting to a time that is within the stored trend period, previously
stored trends are not erased.
Changing the day, month, or year setting does not affect the stored patient trends.
BR
TEMP
100.4
98.6
1.8
97
( 85)
mmHg
(
M
)
%
F
58
MCO
2
S
p
O
2
mmHg
38
Br/m
D
NEXT
T1
T2
ΔT
SERVICE
TIME DAY
UP DOWN ENTER
H:MIN:S
15:55:10
MO/DA/YR
10/21/99
12
WARNING Changing the hour/minute/second setting for the monitor in the
Time/Day window can cause the monitor to erase previously stored patient trend
data.