Vital Signs Monitor 300 Series Directions for use Software version 1.
ii Welch Allyn Vital Signs Monitor 300 Series © 2014 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn.
iii Contents 1 - General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Product overview . . . . . . . . . .
iv Contents Welch Allyn Vital Signs Monitor 300 Series 7 - Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Battery operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Monitor specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Factory default settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 1 General information About this manual This manual contains information about the Welch Allyn® Vital Signs Monitor 300 Series monitor.
2 Chapter 1 General information Welch Allyn Vital Signs Monitor 300 Series Symbols The symbols illustrated on the following pages appear on the monitor or in this document. Table 1. Symbols: Certification and Operation This device has been tested and certified by Urgent alarm notification (output to Nurse Call the Canadian Standards Association system) International to comply with applicable U.S. and Canadian medical safety standards.
Directions for Use Chapter 1 General information 3 The functions of the monitor front panel controls illustrated here are described in detail elsewhere in this document. Table 5. Front Panel Controls Set alarm limits Power on/off Silence alarms Print patient data Scroll up/down Scroll forward/back Increase/decrease value Review patient data (The scroll icon appears as these two arrows in the documentation.
4 Chapter 1 General information Welch Allyn Vital Signs Monitor 300 Series Caution Anyone connecting additional equipment to the signal input part or signal output part of this monitor configures a medical system and is responsible for verifying that the system complies with the requirements of the system standard IEC 60601-1-1. Changes or modifications not expressly approved by Welch Allyn could void the purchaser’s authority to operate the equipment.
Directions for Use Chapter 1 General information 5 WARNING Do not connect more than one patient to a monitor. WARNING If the monitor is dropped or damaged, it must be thoroughly tested by a qualified service person before it is returned to service. WARNING Periodically check all cords and cables for damage, wear, or fraying; replace as needed. WARNING The monitor contains no operator-serviceable parts, other than the replaceable paper roll.
6 Chapter 1 General information Welch Allyn Vital Signs Monitor 300 Series Displays, indicators, controls, and connections This section describes the measurement displays, status indicators, function controls, and connections of the monitor. Numeric measurement and message displays SYS, DIA, and SpO2. Displays systolic and diastolic blood pressure and SpO2, or related alarm thresholds and error codes. (See “Error codes” on page 53.
Directions for Use Chapter 1 General information 7 Status indicators Temperature Units NIBP Measurement Units kilopascals kPa millimeters of mercury mmHg Pulse Amplitude Indicator ºF ºC degrees Fahrenheit degrees Celsius Temperature Type M monitored Patient type neonate alarms not silenced pediatric adult alarms silenced Battery Status power is off power is on charged ——— charging - - - - discharged
8 Chapter 1 General information Welch Allyn Vital Signs Monitor 300 Series Function controls Print Menu Up/Down Power On/Off Power is on Power is off Review Data Set NIBP Interval Start/Stop NIBP (AUTO button) Set Alarm Limits Silence Alarms Alarms Silenced Connections Nurse Call Cable Connector SpO2 30V x.
9 2 Setup This chapter describes the set-up procedures for patient monitoring. Connections Use the procedures described below to connect components to the monitor. Connecting AC power The monitor operates on DC power, supplied by either the internal battery or the AC power adapter. (For information on the battery, refer to “Battery operation” on page 65 and “Electrical” on page 68.) When the AC power adapter is connected, it simultaneously powers the monitor and charges the internal battery.
10 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series To use the AC power adapter: 1. Plug the power adapter into the AC power source. 2. Plug the power adapter connector into the monitor DC port. AC Power Adapter Port Use the AC power adapter to fully charge the battery before using the monitor. (This can take up to 12 hours.) Caution Fully charge the battery before using the monitor for the first time. Failure to do so will result in poor battery performance and reduced battery life.
Directions for Use Chapter 2 Setup 11 Connecting the NIBP cuff hose Attach the hose to the monitor and the cuff as follows, referring to the illustration below: 1. Screw the hose connector onto the NIBP connector on the monitor. 2. Connect the monitor hose connector to the mating connector on the cuff. Threaded NIBP Hose Connector For information on NIBP measurements, see “Patient monitoring” on page 29.
12 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Connecting the temperature probe cable Follow these steps to connect the temperature probe cable to the monitor. 1. Locate the temperature probe connector port on the back of the monitor. 2. Holding the temperature probe cable connector with the spring tab on the right, carefully insert it into the monitor temperature probe connector port. The spring tab clicks out when the connector halves are fully and correctly mated. 3.
Directions for Use Chapter 2 Setup 13 Connecting and disconnecting the SpO2 sensor cable To connect the SpO2 sensor cable: 1. Locate the SpO2 sensor cable connector (labeled SpO2) on the side of the monitor. 2. Note the hole patterns of the connector halves, and align the cable connector accordingly. 3. Carefully insert the SpO2 cable connector into the SpO2 monitor connector.
14 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series To disconnect the SpO2 cable, refer to the instructions shown in the illustration below. Note Always grasp the cable by the connector shoulder. Do not pull on the cable itself. Thumb and forefinger squeeze the tabs in the shoulder of the connector cable to release the cable connector. For information on SpO2 measurements, see “Patient monitoring” on page 29.
Directions for Use Chapter 2 Setup 15 Power on, power-on self-test, and power off When the battery is charged, press to turn on the monitor. The monitor runs a diagnostic self-test each time it powers up. • If all tested functions are working normally, the various windows briefly display start-up values (‘8’ and ‘- -’) and a short tone sounds twice. Caution Users should check for audible alarm function every time the VSM 300 is used.
16 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Configuring operating parameters You can change several monitor operating parameters. When changed, these settings become the default power-up settings. How to use the menu system The monitor menu system contains three sets of menus—settings, configuration, and service. Settings menu Access the settings menu by pressing the menu button Then press repeatedly to reach the setting of interest. while in normal operation.
Directions for Use Chapter 2 Setup 17 Use the settings menu to select and set the following parameters: Patient Type Target Pressure Temp Modes Pulse Tone Volume Neonate Term birth through 28 days, or up to 44 gestational weeks Pediatric 29 days through 12 years Adult 13 years and older The initial cuff inflation pressure (set individually for each patient type) Fahrenheit Predictive ºF ºF M Fahrenheit Monitored Celsius Predictive ºC ºC M Celsius Monitored From 0 (silent) to 5 (loudest) To chan
18 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Configuration menu The configuration menu is accessed by pressing and keeping it depressed for three seconds. You then press repeatedly until you reach the setting of interest. Configuration Menu 128 71 54 98 Set Time and Date 37.
Directions for Use Chapter 2 Setup 19 Use the configuration menu to select and set the following parameters: Time and Date hour minute year month day MAP Measurement Enabled Disabled Blood Pressure Measurement Units mmHg (millimeters of mercury) kPa (kilopascals) Print Mode Batch Stream To change a configuration parameter: 1. Select the parameter as indicated above. 2. Change the value by pressing or . 3.
20 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Changing the time and date Follow these steps to change the time and date settings of the monitor internal clock. 1. Press and hold for 3 seconds. SET HOUR XX appears in the message window. /min mmHg SYS SpO2 % ºF DIA SET HOUR 00 2. Press or as needed to change XX to the current hour. 3. Press once to set the hours and change the display to SET MINUTE XX. 4. Press or 5.
Directions for Use Chapter 2 Setup 21 11. To save the displayed time and date settings, either do nothing for 10 seconds or press any button other than or . If you press a function button (such as ), the monitor returns to normal operation with that function ( ) activated. /min 0 mmHg SYS SpO2 % 0 ºF DIA 0 09:24:17 You cannot change the date and time while memory contains stored vital-signs data.
22 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Changing the patient type The age range for each patient type is defined as follows: Neonatal Term birth through 28 days, or up to 44 gestational weeks Pediatric 29 days through 12 years Adult 13 years and older Default setting: ADULT. Follow these steps to change the patient type setting. 1. Press . The current patient type ( , , or ) appears below the DIA window, and NEONATE, PEDIATRIC, or ADULT appears in the message window. 2.
Directions for Use Chapter 2 Setup 23 MAP measurement enable and disable Default setting: MAP ENABLED for neonate; MAP DISABLED for adult and pediatric. 1. Depress for 3 seconds. SET HOUR XX appears in the message window. 2. Press repeatedly until MAP ENABLED or MAP DISABLED appears in the display window. /min mmHg SYS SpO2 % ºF DIA MAP DISABLED 3. Press Note or to enable or disable MAP measurement.
24 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Changing the NIBP measurement units Default setting: mmHg. To change the NIBP measurement units: 1. Depress for 3 seconds. SET HOUR XX appears in the message window. 2. Press repeatedly until BP Units: mmHg or BP Units: kPa appears in the display window. /min mmHg SYS SpO2 % ºF DIA BP Units: mmHg 3. Press or as needed to display the desired NIBP measurement units. 4.
Directions for Use Chapter 2 Setup 25 Changing temperature type and measurement units Default setting: F (Fahrenheit predictive). To change the temperature type and the temperature measurement units: 1. With the monitor on, press repeatedly until TEMP MODE appears in the display window. One or two green LEDs to the right of the temperature window illuminate to indicate the selected temperature type. /min mmHg SYS SpO2 % ºC DIA M TEMP MODE 2.
26 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series Changing the volume of the pulse tone Default setting: 03. The pulse tone can be set from level 00 (volume off) to level 05 (volume on full). To adjust the volume of the SpO2 pulse tone, do the following: 1. Press repeatedly until VOLUME XX appears in the display window and the pulse tone sounds continuously. /min kPa SYS SpO2 % ºC DIA M VOLUME 03 2. Press or to raise or lower the volume level.
Directions for Use Chapter 2 Setup 27 Selecting stream or batch printing Default setting: BATCH. For monitors configured with the optional thermal printer: 1. Press and hold for three seconds. 2. Press until the message window reads PRINT: BATCH or PRINT: STREAM. /min kPa SYS SpO2 % ºC DIA M PRINT: STREAM 3. Press or to alternate between PRINT: BATCH and PRINT: STREAM display. 4.
28 Chapter 2 Setup Welch Allyn Vital Signs Monitor 300 Series
29 3 Patient monitoring Monitoring blood pressure WARNING To ensure safe and accurate NIBP measurements, use only cuffs and hoses approved by or supplied by Welch Allyn. WARNING Never use an adult or pediatric monitor setting or cuff for an NIBP measurement on a neonatal patient. Adult and pediatric inflation limits can be excessive for neonatal patients, even if a neonatal cuff is used. WARNING NIBP readings may be inaccurate for patients experiencing moderate to severe arrhythmia.
30 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series Changing the target pressure Follow these steps to change the target pressure (default initial pressure for cuff inflation) for the current patient type: 1. Press until the message window displays TARGET PRESSURE. The SYS window displays the current setting for initial inflation pressure. 2. Press or to raise or lower the preset pressure value to the target level.
Directions for Use Chapter 3 Patient monitoring Cuff Size Circumference Circumference (inches) (centimeters) Cuff Size 31 Circumference Circumference (inches) (centimeters) Neonate #1 1.3 - 2.2 3.3 - 5.6 Small Child 4.9 - 6.6 12.4 - 16.8 Neonate #2 1.6 - 2.8 4.2 - 7.1 Child 6.2 - 8.4 15.8 - 21.3 Neonate #3 2.1 - 3.6 5.4 - 9.1 Small Adult 7.9 - 10.6 20.0 - 27.0 Neonate #4 2.4 - 4.6 6.9 - 11.7 Adult 10.0 - 13.5 25.3 - 34.4 Neonate #5 3.5 - 5.9 8.9 - 15.0 Large Adult 12.
32 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series 1. Attach the cuff to the patient’s arm. 2. Press . • The monitor inflates the cuff. • The SYS window dynamically displays the current cuff pressure. Note If the message ‘CAL’ appears in the message window when you attempt to start an NIBP cycle, it means that the NIBP measurement system is self-calibrating to a zero baseline and is temporarily unavailable (for up to 30 seconds).
Directions for Use Chapter 3 Patient monitoring 33 To set up an automatic NIBP measurement, do the following: 1. Attach the cuff to the patient’s arm. 2. Press to set the measurement interval. The two dashes (- -) in the message window indicate that automatic measurement is turned off. /min mmHg SYS SpO2 % ºC DIA M INTERVAL -- 3. To set an interval, press or to cycle through the options, which include - -, ST, and a range of intervals: 1, 3, 4, 5, 10, 15, 30, 45, 60, 90, 120, and 240 minutes.
34 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series • the next cycle begins, if the monitor is still in automatic NIBP mode • two minutes pass • is pressed again If the first cycle does not produce a measurement, the monitor retries the measurement using a target pressure calculated from the results of the previous cycle. Note The automatic NIBP cycles continue until one of the following occurs: • The monitor reaches the 5-minute limit for a STAT measurement.
Directions for Use Chapter 3 Patient monitoring 35 For example: If the monitor is set to Neonatal and you set MAP Disabled, MAP Disabled becomes the default setting for neonatal patients until you change the enabled/disabled setting again. Enabling and disabling MAP measurement See “MAP measurement enable and disable” on page 23.
36 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series Monitoring SpO2 Warnings and cautions — SpO2 WARNING Disconnect the SpO2 cable from the monitor before defibrillating the patient. WARNING Always follow the manufacturer’s instructions for care and use of the SpO2 sensor.
Directions for Use Chapter 3 Patient monitoring 37 Caution Some sensors might not work with some patients. If, after 20 seconds, a properly functioning sensor fails to discern a pulse, do the following: 1. Adjust or reposition the sensor. If the failure continues: 2. Use a different type of sensor. SpO2 monitoring procedure 1. Verify that the SpO2 sensor cable is connected to the monitor. 2. Attach the SpO2 finger clip sensor to the end of the patient’s index finger, as shown below.
38 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series Caution Users should check for audible pulse tones in conjunction with the SpO2 function. If these tones do not sound, the audio has failed. Remove the device from service and contact Welch Allyn.
Directions for Use Chapter 3 Patient monitoring 39 Monitoring temperature Warnings and cautions — temperature WARNING To ensure patient safety and to obtain accurate and reliable temperature results, read this section thoroughly before using the temperature instrument. WARNING Always put a single-use probe tip cover on the probe tip before taking a temperature measurement. Failure to use a probe tip cover can cause patient discomfort, patient cross-contamination, and erroneous temperature readings.
40 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series Setting the temperature measurement type The monitor, if configured with the temperature option, can provide both predictive and monitored temperature measurements. A predictive measurement is a one-time measurement that takes only a few seconds. It results in a single temperature reading which is displayed at the end of the brief measurement period.
Directions for Use Chapter 3 Patient monitoring 41 Ejecting a used probe cover Do not touch the used probe cover. 1. Position the probe over an appropriate disposal receptacle. 2. While holding the probe securely, push the probe cover ejector button (blue or red) to remove the probe cover into the disposal receptacle. Predictive temperature measurement Note Verify that the temperature measurement type is set to predictive. (The display is either ºF or ºC ; the letter ‘M’ is not illuminated.
42 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series Sublingual pockets 4. Have the patient close his/her lips around the probe. Caution If the patient bites the probe, the probe can be damaged. 5. Hold the probe in place to assure continuous contact with the oral tissue until the measurement is complete. Rotating segments appear in the temperature window, indicating that the measurement is in progress.
Directions for Use Chapter 3 Patient monitoring 43 WARNING If the probe becomes contaminated, follow the instructions under “Thermometer and probe cleaning procedure” on page 46. Axillary predictive When used correctly, the monitor produces an accurate axillary temperature measurement in less than 15 seconds for adults and in less than 13 seconds for pediatric patients. Note For axillary temperatures, use only the oral probe (blue ejection button) and the blue probe well.
44 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series 7. Eject the probe cover by pressing the ejection button; hygienically dispose of the probe cover. 8. Return the probe to the probe well. If the monitor cannot make a predicted measurement within 60 seconds, it switches to making a monitored temperature measurement. (See “Monitored temperature measurement” on page 45.) Caution Do not monitor temperature continuously for more than 5 minutes.
Directions for Use Chapter 3 Patient monitoring 45 7. Eject the probe cover by pressing the ejection button, and hygienically dispose of it. 8. Return the probe to the probe well. WARNING If the probe becomes contaminated, follow the instructions under “Thermometer and probe cleaning procedure” on page 46. Monitored temperature measurement Caution Do not monitor temperature continuously for more than 5 minutes. Note Verify that the temperature measurement type is set to monitored.
46 Chapter 3 Patient monitoring Welch Allyn Vital Signs Monitor 300 Series Thermometer and probe cleaning procedure 1. Wipe the thermometer regularly with a cloth dampened with warm water and a mild detergent solution. 2. Occasionally clean the thermometer and probe as necessary with either a 70% isopropyl alcohol or a 10% solution of chlorine bleach. Caution Do not immerse or soak the thermometer or probe in any type of fluid.
47 4 Alarms and alerts WARNING If you turn off any alarm limits while responding to an alarm, verify alarm limits before you resume patient monitoring. Responding to a patient alarm WARNING If a patient alarm and an equipment alert occur at the same time, take care of the patient alarm first. A patient alarm occurs when a vital-sign measurement falls outside of programmed limits.
48 Chapter 4 Alarms and alerts Welch Allyn Vital Signs Monitor 300 Series Recoverable SpO2 alert—escalated An SpO2 equipment alert is always escalated immediately to the level of a patient alarm if it occurs when both of the following conditions exist: • SpO2 monitoring has begun and the monitor has recorded an SpO2 measurement • An SpO2 or Pulse Rate alarm limit has been set See “Responding to a patient alarm” on page 47.
Directions for Use Chapter 4 Alarms and alerts 49 Alarm indicators The monitor alarm indicators are as follows: Event Audible indicator Visual indicator Patient alarm Three short tones in quick succession, followed by a short silence, and then two short tones in quick succession, followed by a long silence; repeated until action is taken. Flashing display of the violating value. For a MAP violation, the monitor repeatedly flashes the MAP numerics in the message display.
50 Chapter 4 Alarms and alerts Note Welch Allyn Vital Signs Monitor 300 Series For patient safety, all alarms are reset to the factory default levels whenever the patient type is changed. This means that you must either accept the default alarm limits or set new limits every time you change patient type ( , , or ). The ‘high’ alarm for any vital sign is always higher than the ‘low’ alarm for the same vital sign.
Directions for Use Chapter 4 Alarms and alerts 51 Alarm limits for vital signs Vital sign Low limit (default) High limit (default) Resolution Neonatal 30 - 110 (50) 35 - 115 (100) 5 mmHg (0.5 kPa) Pediatric 35 - 150 (75) 40 - 155 (145) 5 mmHg (0.5 kPa) Adult 35 - 250 (75) 40 - 255 (220) 5 mmHg (0.5 kPa) Neonatal 15 - 95 (30) 20 - 100 (70) 5 mmHg (0.5 kPa) Pediatric 20 - 120 (35) 25 - 125 (100) 5 mmHg (0.5 kPa) Adult 25 - 225 (35) 30 - 230 (110) 5 mmHg (0.
52 Chapter 4 Alarms and alerts Welch Allyn Vital Signs Monitor 300 Series To connect the monitor to a Nurse Call system, you must have a cable (Welch Allyn part number 008-0634-XX or equivalent) that has been adapted to your Nurse Call system. This cable can be built or customized by a biomedical engineering department or other appropriate technical resource, using the specification described on “Nurse call” on page 73.
Directions for Use Chapter 4 Alarms and alerts Error codes Error code Description Blood pressure C01 NIBP cycle cancelled by operator. C02 Unable to calibrate. Either the cuff is being moved excessively during the calibration or the NIBP unit is not working. C03 Inflation too rapid. Check for hose constriction. Verify patient type setting. C04 Inflation too slow. Check for hose leak. Verify patient type setting. C05 Excessive noise or air leak.
54 Chapter 4 Alarms and alerts Welch Allyn Vital Signs Monitor 300 Series
55 5 Reviewing patient data You can review stored patient data either by viewing it on the monitor or by printing it. Displaying stored patient data 1. Press to display the most recent stored set of patient vital-signs data. • The monitor interrupts the dynamic display of any current vital-sign measurement. • The message READING appears in the message window, with a data-set sequence number and the time of the displayed reading.
56 Chapter 5 Reviewing patient data Welch Allyn Vital Signs Monitor 300 Series Batch printing After a patient has been continuously monitored over some period, a clinician can use the monitor’s batch printing capability to print all of the measurements that were stored in the monitor’s memory over the monitoring period.
Directions for Use Chapter 5 Reviewing patient data 57 Header Welch Allyn® | | | | Vital Signs Monitor | | Patient information | | | | | | | | Patient Name:___________________ | Patient ID:_____________________ | Physician:______________________ | Procedure:______________________ | Comments:_______________________ | ________________________________ | ________________________________ | | Patient data The contents of the Patient Data block reflect the configuration of the monitor; that is, with or wit
58 Chapter 5 Reviewing patient data Welch Allyn Vital Signs Monitor 300 Series Date | DD-MMM-YYYY | Time The recording time is printed for each record, in the form hh:mm (using a 24-hour clock), at the beginning of the first line of the record.
Directions for Use Chapter 5 Reviewing patient data 59 Pulse rate The pulse rate notation does not include error information. SpO2 SpO2 notation can include error information: C7 Malfunctioning sensor. -- Sensor error. NIBP NIBP notation can include error information: C01 NIBP cycle cancelled by operator. C02 Unable to calibrate. C03 Cuff inflation to rapid. C04 Excessive cuff inflation time. C05 Excessive noise or air leak. C06 Measurement out of range. C10 Cuff overpressure.
60 Chapter 5 Reviewing patient data Welch Allyn Vital Signs Monitor 300 Series Erasing patient data All patient vital-sign data is erased when the monitor is powered off or when you change the time and date settings. You can also erase data at any time during normal monitor operation. Erasing data before changing the date and time If you attempt to change the date and time, the monitor prompts you to confirm that you also want to erase all stored patient data.
Directions for Use Chapter 5 Reviewing patient data 61 Replacing the printer paper supply Note Use only the thermal paper (part number 7052-25) supplied by Welch Allyn. The use of any other paper can result in poor printer performance. To replace the printer paper supply, follow these steps. 1. Press to open the printer door. 2. Insert a new roll of paper. Note The paper roll must be installed as shown in the illustration below.
62 Chapter 5 Reviewing patient data Welch Allyn Vital Signs Monitor 300 Series
63 6 Operator maintenance Cleaning Wipe the monitor with a cloth slightly dampened with warm water and a mild detergent or appropriately diluted, nonstaining disinfectant solution. Caution Never wet the monitor or immerse it in fluid of any kind. Never allow water to enter any connector ports. Caution Never immerse any monitor components, such as the NIBP hose or cuff, the temperature probe, or the SpO2 sensor.
64 Chapter 6 Operator maintenance Welch Allyn Vital Signs Monitor 300 Series Recycling monitor components Within the EU Do not dispose of this product as ‘unsorted municipal waste’. Prepare it for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). Note If the monitor or the battery is contaminated, this directive does not apply.
65 7 Reference Battery operation Battery low warning When the battery charge is low, a warning tone sounds and the Low Battery indicator flashes continuously.
66 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series Monitor specifications The monitor is an FDA Class II non-critical device and, according to Council Directive 93/ 42/EEG, Annex IX, a Class IIB device. Performance Characteristic Recovery from Defibrillation Discharge Cuff Pressure Range Initial Cuff Inflation Factory Default Specification Notes General Per IEC 60606-2-30:1999(E) The monitor returns to normal function within 1 minute after the discharge of a cardiac defibrillator.
Directions for Use Chapter 7 Reference Characteristic Specification Notes Temperature Temperature Measurement Range Temperature Calibration Accuracy Temperature Determination Time 80 F to 110 F 26.7 C to 43.3 C ± 0.2 F (± 0.
68 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series Characteristic Specification Support Self-supporting on rubber feet Mountable (with available accessories) to the following: IV pole Mobile stand Wall Portability Recessed carry handle Mounted on a pole or mobile stand, can be rolled from patient to patient Electrical Power requirements The medical-rated isolation AC power adapter connected to the AC supply must supply 8 volts DC and 0.74 amperes.
Directions for Use Chapter 7 Reference Electrical rating descriptions Power option Specification 1 Input: 120VAC 60Hz , 130mA Output: 8VDC, 750mA 2 Input: 240VAC 50Hz , 65mA Output: 8VDC, 750mA Input: 230VAC 60Hz , 65mA Output: 74VDC, 750mA 4 Input: 240VAC 50Hz , 65mA Output: 8VDC, 750mA Input: 230VAC 60Hz , 65mA Output: 74VDC, 750mA 6 Input: 240VAC 50Hz , 65mA Output: 8VDC, 750mA Input: 230VAC 60Hz , 65mA Output: 74VDC, 750mA 7 Input: 220VAC 50-60Hz , 70mA Output: 8VDC, 750mA 69
70 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series Environmental EMC compliance The monitor complies with all applicable and required standards for electromagnetic interference. • It does not normally affect nearby equipment and devices. • It is not normally affected by nearby equipment and devices. • It is safe to operate the monitor in the presence of high-frequency surgical equipment.
Directions for Use Chapter 7 Reference 71 Guidance and manufacturer’s declaration—electromagnetic immunity The Vital Signs Monitor 300 Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
72 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series Guidance and manufacturer’s declaration—electromagnetic immunity The VSM 300 Series monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the monitor should assure that it is used in such an environment.
Directions for Use Chapter 7 Reference 73 Recommended separation distances between portable and mobile RF communications equipment and the Vital Signs Monitor 300 Series monitor The Vital Signs Monitor 300 Series monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
74 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series Nurse call Switch current 1A maximum Switch voltage 30V ac/dc maximum Isolation 1500 Vrms Alarm relay Energized during patient alarm Cable Welch Allyn part number 008-0634-XX, customized to connect to the local Nurse Call system. Cable properties: • 10 feet (3 meters) long. • Not shielded. • One end not terminated (to be customized on site).
Directions for Use Chapter 7 Reference Factory default settings Function Blood pressure measurement units Patient type Automatic NIBP Interval MAP (Adult and Pediatric) MAP (Neonate) NIBP Adult High Systolic Alarm NIBP Adult Low Systolic Alarm NIBP Adult High Diastolic Alarm NIBP Adult Low Diastolic Alarm NIBP Adult High MAP Alarm NIBP Adult Low MAP Alarm NIBP Pediatric High Systolic Alarm NIBP Pediatric Low Systolic Alarm NIBP Pediatric High Diastolic Alarm NIBP Pediatric Low Diastolic Alarm NIBP Pediatr
76 Chapter 7 Reference Welch Allyn Vital Signs Monitor 300 Series
75 Limited warranty This product is sold by Welch Allyn under the warranties set forth in the following paragraphs. These warranties are extended only to the end-user with respect to the original purchase of this product directly from Welch Allyn or from Welch Allyn’s authorized distributors.
76 Limited warranty Welch Allyn Vital Signs Monitor 300 Series
77 Index A AC power connecting, 9 adult see also patient type defined, 22 icon, 3 alarm high and low, 50 MAP, 33, 50 patient, 47 range limits, 50 responding to, 47 setting, 49 tone, 47 alert responding to, 47 autoclave, 5 axillary temperature, 43 B batch printing, 27, 56 battery charge before using, 10 charging, 9 failure, 65 low power indicator, 10 blood pressure automatic, 32 manual, 31 monitoring, 29 buttons description, 8 charging the battery, 9 classification, 66 cleaning, 63 configuration, 16 conne
78 Index K kPa units, selecting, 24 M maintenance, 63 MAP alarm, 33, 50 measurement data, reviewing, 55 units, setting, 24 mmHg units, selecting, 24 monitor configuration, 16 setup, 16 monitored temperature, 40 monitoring blood pressure, 29 pulse rate, 35 SpO2, 36 temperature, 39 N neonatal see also patient type defined, 22 icon, 3 NIBP, 29 automatic, 32 cycle, ending, 34 manual, 31 preparation, 29 STAT, 35 nurse call connector location, 51 Welch Allyn Vital Signs Monitor 300 Series and MAP enable/dis
Directions for Use blood pressure, 29 pulse rate, 35 SpO2, 36 temperature, 39 W warnings general, 4 SpO2, 36 Index 79
80 Index Welch Allyn Vital Signs Monitor 300 Series