When a Propaq CS monitor in Adult or Pediatric Mode is connected to an Acuity System, the
audible alarms at the bedside Propaq CS monitor can be delayed up to 4 minutes and 15
seconds. The delay time is selected in Acuity software at the time of Acuity installation. Visual
alarm indications are not delayed.
Disconnect the Hardwired Monitor from the Acuity System
1.
To permanently disconnect the Propaq CS monitor from the Acuity
network, press the
NET OFF
key on the upper left corner of the monitor
display. (From the Main Menu you can also press
SETUP,
ACUITY
,
NET
OFF
.)
2.
Within 15 seconds, disconnect the Acuity network cable from the
Propaq CS monitor side panel and the bedside jack. If the patient will
no longer be monitored with this monitor, turn off the monitor to erase
trend information.
If you want to temporarily disconnect the Propaq CS monitor from the
Acuity network and reconnect the same patient to the Acuity network later,
see the
Acuity System Reference Guide
.
Note
If the Propaq CS has the wireless option, be aware that after you disconnect the hardwire
connection and disconnect the Acuity network cable, the Wireless Propaq CS will attempt to
establish a wireless connection to Acuity as long as the power is on.