User's Manual Addendum
Table Of Contents
- UPDATE
- Contents
- NEW ROOM
- 1. If the Propaq CS monitor has already been connected to the patient, save the patient’s Trends ...
- 2. If the monitor is not already connected to the patient, attach leads and sensors to the patien...
- 3. Plug in the Acuity network cable to the Acuity network jack on the monitor side panel as shown...
- 4. Connect the ac power adapter to the monitor and the wall outlet to charge the battery. Check t...
- 5. When the monitor completes the connection to Acuity, it displays the message ON NETWORK (or al...
- 6. If alarm limits have not been set, do so at the monitor or at the Acuity Central Station using...
- 1. To permanently disconnect the Propaq CS monitor from the Acuity network, press the NETOFF key...
- 2. Within 15 seconds, disconnect the Acuity network cable from the Propaq CS monitor side panel a...
- 1. Press the NETOFF key on the upper left corner of the monitor display to disconnect from Acuit...
- 2. Within 15 seconds unplug the Acuity network cable from the Wireless Propaq CS Acuity connector...
- 3. After the network connection is established, select the Acuity Unit, patient, and room from th...
- 1. Make sure the Acuity network cable is not plugged into the Acuity connector and an Acuity beds...
- 2. Confirm that the Wireless Propaq CS displays the flashing message SEARCHING in the upper left ...
- 3. Confirm that within one minute the monitor displays these alternating messages:
- 4. After the network connection is established, the monitor may prompt you to select an Acuity Un...
- 5. The Wireless Propaq CS displays a list of possible patients. If your patient has been pre-admi...
- 6. Within a few seconds the Wireless Propaq CS displays a list of unassigned rooms.
- 7. Customize alarm limits for your patient if needed. Refer to the Propaq CS Directions For Use.
- 1. A DROPOUT equipment alert occurs at the Acuity Central Station. Acknowledge the alert at Acuity.
- 2. An equipment alert occurs at the Wireless Propaq CS with this message:
- 1. Press the NETOFF key on the upper left corner of the monitor display. (From the Main Menu you...
- 2. Disconnect the leads and sensors from the patient and turn off the power to the monitor.
- 1. Turn on power to the Wireless Propaq CS and confirm that the monitor displays the powerup screen.
- 2. The Wireless Propaq CS will then present a series of menus and messages requesting you to prov...
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEW ROOM.
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, TRANSFER. Within a few seconds the Wi...
- 2. Scroll up or down to highlight the new unit, then press Select.
- 3. After the patient is assigned to the new unit, the Wireless Propaq CS displays a list of unass...
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEWPATIENT.
- 1. Plug the Acuity network cable into the Wireless Propaq CS Acuity connector and an Acuity bedsi...
- NEW ROOM
14 Welch Allyn
Monitor a Patient Out of Range of Acuity
While out of range of Acuity, the Wireless Propaq CS continues to monitor
the patient and provide local alarms or alerts at the Propaq CS as needed.
When the patient with the Wireless Propaq CS goes out of range of Acuity, do
the following:
1.
A DROPOUT equipment alert occurs at the Acuity Central Station.
Acknowledge the alert at Acuity.
2.
An equipment alert occurs at the Wireless Propaq CS with this
message:
NETWORK FAULT
CHECK ACUITY/NETWORK CONNECTION
Press a key on the Propaq CS to acknowledge the alert and silence the
alert tones (if tones are enabled).
When the patient with the Wireless Propaq CS returns within range of
Acuity, the Propaq CS automatically reconnects to Acuity. No clinician
intervention is required.