User's Manual Addendum
Table Of Contents
- UPDATE
- Contents
- NEW ROOM
- 1. If the Propaq CS monitor has already been connected to the patient, save the patient’s Trends ...
- 2. If the monitor is not already connected to the patient, attach leads and sensors to the patien...
- 3. Plug in the Acuity network cable to the Acuity network jack on the monitor side panel as shown...
- 4. Connect the ac power adapter to the monitor and the wall outlet to charge the battery. Check t...
- 5. When the monitor completes the connection to Acuity, it displays the message ON NETWORK (or al...
- 6. If alarm limits have not been set, do so at the monitor or at the Acuity Central Station using...
- 1. To permanently disconnect the Propaq CS monitor from the Acuity network, press the NETOFF key...
- 2. Within 15 seconds, disconnect the Acuity network cable from the Propaq CS monitor side panel a...
- 1. Press the NETOFF key on the upper left corner of the monitor display to disconnect from Acuit...
- 2. Within 15 seconds unplug the Acuity network cable from the Wireless Propaq CS Acuity connector...
- 3. After the network connection is established, select the Acuity Unit, patient, and room from th...
- 1. Make sure the Acuity network cable is not plugged into the Acuity connector and an Acuity beds...
- 2. Confirm that the Wireless Propaq CS displays the flashing message SEARCHING in the upper left ...
- 3. Confirm that within one minute the monitor displays these alternating messages:
- 4. After the network connection is established, the monitor may prompt you to select an Acuity Un...
- 5. The Wireless Propaq CS displays a list of possible patients. If your patient has been pre-admi...
- 6. Within a few seconds the Wireless Propaq CS displays a list of unassigned rooms.
- 7. Customize alarm limits for your patient if needed. Refer to the Propaq CS Directions For Use.
- 1. A DROPOUT equipment alert occurs at the Acuity Central Station. Acknowledge the alert at Acuity.
- 2. An equipment alert occurs at the Wireless Propaq CS with this message:
- 1. Press the NETOFF key on the upper left corner of the monitor display. (From the Main Menu you...
- 2. Disconnect the leads and sensors from the patient and turn off the power to the monitor.
- 1. Turn on power to the Wireless Propaq CS and confirm that the monitor displays the powerup screen.
- 2. The Wireless Propaq CS will then present a series of menus and messages requesting you to prov...
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEW ROOM.
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, TRANSFER. Within a few seconds the Wi...
- 2. Scroll up or down to highlight the new unit, then press Select.
- 3. After the patient is assigned to the new unit, the Wireless Propaq CS displays a list of unass...
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEWPATIENT.
- 1. Plug the Acuity network cable into the Wireless Propaq CS Acuity connector and an Acuity bedsi...
- NEW ROOM
19 Welch Allyn
Reassign the Wireless Propaq CS to a New Patient
If you want to discontinue monitoring a patient and connect the Wireless
Propaq CS to a new patient, follow these steps.
1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT,
NEW PATIENT.
The monitor then presents a series of menus and messages requesting
you to provide information about the connection and patient. The
actual screens presented depend on how the Acuity System is
configured.
Provide the information as requested. This may include:
• Select an Acuity unit.
• Select a patient from the patient list. (After you select a new patient,
all monitor operating settings are reset to the Acuity System default
powerup settings.)
• Select a patient room from the room list.
If you do not select the patient name or room while connecting the patient,
you will need to do that later at the Acuity Central Station. See the Acuity
System Reference Guide or Acuity Directions For Use for more information.
Switch from Wireless to Hardwired Monitoring
1. Plug the Acuity network cable into the Wireless Propaq CS Acuity
connector and an Acuity bedside network jack.
The Wireless Propaq CS will automatically switch to a hardwired
Acuity network connection. No equipment alerts are generated.