User's Manual Addendum
Table Of Contents
- UPDATE
- Contents
- NEW ROOM
- 1. If the Propaq CS monitor has already been connected to the patient, save the patient’s Trends ...
- 2. If the monitor is not already connected to the patient, attach leads and sensors to the patien...
- 3. Plug in the Acuity network cable to the Acuity network jack on the monitor side panel as shown...
- 4. Connect the ac power adapter to the monitor and the wall outlet to charge the battery. Check t...
- 5. When the monitor completes the connection to Acuity, it displays the message ON NETWORK (or al...
- 6. If alarm limits have not been set, do so at the monitor or at the Acuity Central Station using...
- 1. To permanently disconnect the Propaq CS monitor from the Acuity network, press the NETOFF key...
- 2. Within 15 seconds, disconnect the Acuity network cable from the Propaq CS monitor side panel a...
- 1. Press the NETOFF key on the upper left corner of the monitor display to disconnect from Acuit...
- 2. Within 15 seconds unplug the Acuity network cable from the Wireless Propaq CS Acuity connector...
- 3. After the network connection is established, select the Acuity Unit, patient, and room from th...
- 1. Make sure the Acuity network cable is not plugged into the Acuity connector and an Acuity beds...
- 2. Confirm that the Wireless Propaq CS displays the flashing message SEARCHING in the upper left ...
- 3. Confirm that within one minute the monitor displays these alternating messages:
- 4. After the network connection is established, the monitor may prompt you to select an Acuity Un...
- 5. The Wireless Propaq CS displays a list of possible patients. If your patient has been pre-admi...
- 6. Within a few seconds the Wireless Propaq CS displays a list of unassigned rooms.
- 7. Customize alarm limits for your patient if needed. Refer to the Propaq CS Directions For Use.
- 1. A DROPOUT equipment alert occurs at the Acuity Central Station. Acknowledge the alert at Acuity.
- 2. An equipment alert occurs at the Wireless Propaq CS with this message:
- 1. Press the NETOFF key on the upper left corner of the monitor display. (From the Main Menu you...
- 2. Disconnect the leads and sensors from the patient and turn off the power to the monitor.
- 1. Turn on power to the Wireless Propaq CS and confirm that the monitor displays the powerup screen.
- 2. The Wireless Propaq CS will then present a series of menus and messages requesting you to prov...
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEW ROOM.
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, TRANSFER. Within a few seconds the Wi...
- 2. Scroll up or down to highlight the new unit, then press Select.
- 3. After the patient is assigned to the new unit, the Wireless Propaq CS displays a list of unass...
- 1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEWPATIENT.
- 1. Plug the Acuity network cable into the Wireless Propaq CS Acuity connector and an Acuity bedsi...
- NEW ROOM
Update Propaq CS DFU 3.4X 3
About this Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
General Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Understanding the Propaq CS, Acuity, and the FlexNet
Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
The Acuity Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Hardwired Connection to Acuity
Connect the Monitor to the Acuity System . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Disconnect the Hardwired Monitor from the Acuity System . . . . . . . . . . . . .10
Switch from Hardwired to Wireless Monitoring . . . . . . . . . . . . . . . . . . . . . . .11
Wireless Connection to Acuity . . . . . . . . . . . . . . . . . . . .12
Connect a New Patient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Monitor a Patient Out of Range of Acuity . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Stop Monitoring a Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Reconnect a Recently Monitored Patient . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Reassign a Monitored Patient to a New Room in the Same Unit . . . . . . . . .17
Transfer a Monitored Patient to a New Room in a Different Unit . . . . . . . . .18
Reassign the Wireless Propaq CS to a New Patient . . . . . . . . . . . . . . . . . .19
Switch from Wireless to Hardwired Monitoring . . . . . . . . . . . . . . . . . . . . . . .20
Equipment Alert Messages . . . . . . . . . . . . . . . . . . . . . . .21
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Contents