Propaq CS Directions for Use 105
Reference
Temperature Specifications
Pulse Oximetry (SpO
2
) Specifications
Characteristic Specification
Range 0° to +50°C; 32° to +122°F
Displays T1, T2, and ∆T
Probes Compatible with YSI Series 400 and 700 probes. HP side panel only
compatible with YSI 400 and has HP connector.
Units °C and °F selectable
Channel Accuracy Temperature Range Tolerance
0° to +10°C ±0.2°C
>10° to +50°C ±0.1°C
+32° to +50°F ±0.4°F
>50° to +122°F ±0.2°F
Resolution 0.1°C or °F
Electrosurgery interference
suppression
Included in all models.
Characteristic Specification
Range 0% to 100%
Probe Accuracy
(specified at 28° to 42° C)
Adults:
70% to 100% ±2 digits
50% to 69% ±3 digits
0% to 49% unspecified
Pediatrics:
70% to 100% ±3 digits
Neonates:
70% to 95% ±3 digits
Pulse Rate Range 25 to 250 beats per minute
Pulse Rate Accuracy ±3 beats per minute or 3%, whichever is greater
Sensor Compatibility Compatible only with NELLCOR sensors listed in the Protocol
Systems Products and Accessories booklet.
Electrosurgery interference
suppression
Included in all models.
Alarm Hold-Off Time Period 10 seconds; reset if the sensor reports levels within limits before 10
seconds elapses.