Nurse Call Specifications
Nurse Call Cable Specifications
This cable (Part Number 008-0634-XX) must be customized by a biomedical technician to connect to
the local Nurse Call system.
Propaq CS Monitor Audible Alarm Indications
Alarm Condition Tone
Patient alarm Continuous ON
Apnea alarm ON for 1 second, OFF for 1 second
Equipment alert ON for 1 second, OFF for 4 seconds
Characteristic Specication
Maximum switch current 1 A
Maximum switch voltage 30 V ac/dc
Isolation 1500 Vrms
Alarm relay Energized during apnea alarm or patient alarm
1
1. Pressing the Suspend/Resume Alarm key or SUSPEND suspends the Nurse Call alarm for 90 seconds.
Customized cable
2
(Protocol Systems Part Number
008-0634-XX); see below.
2. Refer to the Protocol Systems Products and Accessories booklet to order the cable.
One end is a 4-pin plug compatible with the monitor
Nurse Call connector; the other end must be
customized to connect to the local Nurse Call system.
Nurse Call Connector
on Monitor
1 Normally Open
4 Not Connected
3 Normally Closed
2 Common (Arm)
30V at 1A MAX
Pin 4
Pin 3
Pin 1
Pin 2
EXTERNAL VIEW
1
2
3
4
Normally Open (Black)
Arm (Red)
Normally Closed (Green)
Not Connected
Nurse Call Cable (Part Number 008-0634-00)