User's Manual

Table Of Contents
Propaq CS Directions for Use 111
Reference
Trends Specifications
Display Specifications
Monitor (Environmental) Specifications
Caution
The monitor may not meet performance specifications if it is not used or stored within these environmental
specifications.
Characteristic Specication
1
1. Assumes SpO
2
and CO
2
functions are present.
Model 242 Parameters NIBP, T1, T2, T, HR (heart rate/pulse rate), SpO
2,
End-tidal CO
2,
Inspired CO
2
, Breath Rate/Resp Rate
Model 244 Parameters NIBP, P1, T1, T2, T, HR (heart rate/pulse rate), SpO
2,
End-tidal CO
2
,
Inspired CO
2
, Breath Rate/Resp Rate
Model 246 Parameters NIBP, P1, P2, T1, T2,
T, HR (heart rate/pulse rate), SpO
2
,
End-tidal
CO
2
,
Inspired CO
2
, Breath Rate/Resp Rate
Duration 5 hours for non-NIBP trends (up to 150 readings)
A maximum of 128 readings (up to 8 hours) for NIBP trends
Resolution All channels except NIBP sample data at 2-minute intervals.
For NIBP trends, a new entry is placed in the table each time an
NIBP determination is made.
Characteristic Specication
Type Color active matrix; TFT (Thin Film Transistor) LCD module
Resolution 640 x 480 pixels; 1 pixel = R + G + B dots
Active Viewing Area 6.73 x 5.10 inches (170.9 x 129.6 mm)
Pixel Pitch 0.0105 inches (0.267 mm)
Viewing Angle U/D 40°, R/L 60° (typical), 10:1 contrast ratio
Contrast Ratio 150:1 (typical); measured in dark room at center of screen
Display Color 18-bit (6 bits per primary color)
Luminance 200 cd/m
2
(typical); measured at saturation point
Response Time 40 ms (maximum); white to black
Characteristic Specication
Operating Temperature 0° to 40° C
Shipping and Storage Temperature -20° to 60° C
Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m)
Operating Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure
1-natural
Shipping and Storage Relative
Humidity
15% to 95%, noncondensing per MIL STD 810E, Procedure
1-natural
Shock 50 g
Vibration, Random 0.02 g
2
/Hz from 10 to 500 Hz, ramping down to 0.002 g
2
/Hz at 2000
Hz. Operating 1 hour per axis, 3 hours per test. Designed to meet
RTCA DO-160D, Category C.
Electromagnetic Compatibility (EMC) EN 60601-1-2: 1993