22 Protocol Systems, Inc.
Change the Current Patient Mode
1. To change the current patient mode, from the Main Menu press SETUP, MORE, CHANGE to access
the Patient Mode window:
2. Based on the patient’s age, press NEONATAL, PEDIATRIC, or ADULT. When the confirmation
window appears, press YES to confirm your selection.
Whenever you change the patient mode, the alarm limit settings, maximum NIBP cuff inflation
pressures, and internal computations are automatically changed to the defaults for that patient mode.
See page 23 for information about preset Factory patient modes or programmable Custom patient
modes.
Note
If you change the patient mode, the CO
2
alarm limits in the new mode might vary slightly from the
originally-programmed CO
2
alarm limits for the new mode. Check the CO
2
alarm limits.
PATIENT MODE
NEONATAL
PEDIATRIC
ADULT
SETUP
SELECT PATIENT MODE BASED ON AGE:
PED : > 44 WEEKS GEST. AGE,
ADULT:
NEO : < 44 WEEKS GEST. AGE
< 9 YEARS
> 9 YEARS
BATTERY: 8.2 VOLTS
PREVIOUS
MENU