User's Manual
Table Of Contents
- Contents
- 1 – General Information
- 2 – Setup
- 3 – Monitoring
- Perform ECG/RESP Monitoring
- Use the Propaq CS Monitor With Pacemaker Patients
- Perform Invasive Blood Pressure (IBP) Monitoring
- Take a Non-Invasive Blood Pressure (NIBP) Reading
- Perform Temperature Monitoring
- Perform SpO2 Monitoring
- Perform Mainstream CO2 Monitoring
- Perform Sidestream CO2 Monitoring
- Set Up the CO2 Display and Alarm Limits
- 4 – Alarms & Alerts
- 5 – Printing & Trends
- 6 – Acuity
- 7 – Defibrillator Synchronization
- 8 – Maintenance
- 9 – Reference
- Index
Propaq CS Directions for Use 23
Setup
Change Powerup Patient Mode or Store Customized Settings
The Propaq CS monitor has standard, preset, default powerup settings and alarm limits for each
patient mode: Adult, Pediatric, and Neonatal. These are “Factory Patient Mode” settings (listed on
page 97).
You can also choose to customize and store programmable powerup settings and alarm limits for each
patient mode. These are “Custom Patient Mode” settings.
The instructions below describe how to change the powerup patient mode and how to select and store
new Custom Patient Mode settings.
Note
When you change patient modes, you also change the alarm limits associated with the new patient mode.
Change the Powerup Patient Mode
1. From the Main Menu, press SETUP, MORE, CHANGE, SETUP to access the Mode Setup window.
2. Press NEXT to highlight the desired Factory or Custom powerup mode, then press POWERUP*
and YES. The new powerup selection is marked by the asterisk.
Changing the powerup mode does not affect the patient mode currently used.
MODE SETUP
BR
TEMP
100.4
98.6
1.8
97
P2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
°F
122
58
MCO2
S
p
O2
mmHg
38
Br/m
S
D
NEXT
T1
T2
∆T
PREVIOUS
MENU
POWERUP* USE
NOW
FACTORY:
SAVE
BATTERY: 9.2 VOLTS
*ADULT PED
CUSTOM
* = PATIENT MODE ON POWERUP
ADULT PED
NEO
NEO
:
12
Asterisk
indicates which
patient mode is
currently
selected for
powerup.