5. If the ZERO menu is not displayed, from the Main Menu press INVASIVE PRESSURE, then ZERO
P1
(or ZERO P2). The word ZEROING appears in the numerics window during zeroing.
If you want to cancel the zeroing process, press CANCEL.
6. Wait for a brief tone to sound and the word ZEROED to appear in the blood pressure
numerics window.
7. Close the transducer's stopcock. The monitor displays the pressure scale and numerics.
8. If the transducer will not zero, the monitor displays the words ZERO REJECTED in the
numerics window. Press CANCEL and try zeroing again beginning at step 4. The monitor does
not display numerics or scales until an acceptable zero reference is established.
You can rezero an IBP transducer at any time after you again open the transducer stopcock to
atmospheric air. If the transducer has already produced pressure readings, rezeroing provides a new
zero reference for the monitor.
If the zero value is not accepted, the monitor continues to use the previous zero reference and
displays numerics and waveforms based on that value.
If the transducer still does not zero, try another transducer or another cable.
Warning
If you press ZERO after an invasive pressure channel has been successfully zeroed and the channel is currently
monitoring a pressure waveform, the message ZERO REJECTED will display in the IBP numerics window. This message
continues to display in place of the valid invasive pressure numerics until you press Home, INVASIVE PRESSURE, and
then CANCEL in the IBP menu. If an IBP alarm occurs while ZERO REJECTED is displayed in place of IBP numerics, the
IBP numerics will not flash to indicate invasive pressure is in alarm.
9. To set up the IBP display, from the Main Menu press INVASIVE PRESSSURE to display the first
IBP menu:
MON
V
BR
TEMP
130
100
70
30
15
0
P
2
P
1
100.4
98.6
1.8
97
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
°F
122
58
MCO2
S
p
O2
mmHg
38
Br/m
12
S
D
RANGE
RESCALE
ZERO P1
ZERO P2
MORE
T1
T2
T
mV
cm
1
LABEL
P
1
FORMAT
P
1
PREVIOUS
MENU
FORMAT
P
2
LABEL
P
2
First IBP
menu
Second IBP
menu
IBP labels are
user-selectable.