Propaq CS Directions for Use 99
Reference
Specications
ECG Specifications
The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and Alarms
specified ANSI/AAMI EC13-1992, except for Standardizing Voltage (section 3.2.9.9). The channel also
meets the American National Standard, Safe Current Limits for Electromedical Apparatus
(ANSI/AAMI ES1-1993).
ECG Specications
Characteristic Specication
Connector AAMI 6 pin or Hewlett-Packard compatible 12-pin style connector
(optional). See illustration on page 100.
Selectable Leads I, II, III, aVR, aVL, aVF, V
Lead Fault Indicator LA, LL, RA, RL, C, multiple
ECG Size (sensitivity) in mV/cm 4, 2, 1, 0.5, 0.2
Display Sweep Speeds 12.5, 25, and 50 mm/s
QRS Tone Volume High, Low, Medium, Off
QRS Tone Frequency 900 Hz for Propaq CS monitor without Expansion Module, 665 Hertz
when equipped with SpO2 but SpO2 not being monitored; variable
pitch with SpO
2
option and SpO
2
being monitored
Bandwidth: MONITOR
EXTENDED
Adult Mode: 0.5 to 40 Hz
Pediatric Mode: 0.5 to 120 Hz
Neonatal Mode: 0.5 to 120 Hz
Adult Mode: 0.05 to 40 Hz
Pediatric Mode: 0.05 to 120 Hz
Neonatal Mode: 0.05 to 120 Hz
(see Real-Time ECG Analog/Deb Sync specication)
Sample Rate 364 Hz
Input Protection Electrosurgery and debrillator protected when used with specied
ECG cables. All models also include electrosurgery interference
suppression.
Lead Fail Sense Current 50 nA dc for active leads
100-200 nA dc for driven lead, depending on number of electrodes
attached
Tall T-wave Rejection Meets AAMI (USA) EC13-1992, section 3.1.2.1.c, for 1.2 mV T-wave
and 1 mV QRS using AAMI test waveform.
Common Mode Rejection <1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source
impedance, input unbalanced, FILTER function OFF
<0.1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source
impedance, input unbalanced, FILTER function ON
Input Impedance >2.5 M differential @ 60 Hz
Input Range (ac) 10 mV peak to peak
Input Range (dc) Up to ±300 mV
System Noise 30 µV peak-to-peak, RTI, with all inputs = 47K in parallel with 0.047
µF.
QRS Detector Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)
Neonatal Amplitude Range: 0.1 to 5.0 mV (RTI)
Neonatal and Pediatric Width Range (Duration):
40 to 120 ms
Adult Width Range (Duration): 70 to 120 ms
Heart Rate Range 25 to 350 beats per minute (measurement)
25 to 300 beats per minute (display)
Heart Rate Meter Response Time Responds to change in heart rate within 5 to 9 seconds depending
on physiological waveform. (As measured per AAMI standard EC
13-1992 clause 4.1.2.1 (f), including 3.1.2.1 parts f. and g.
waveforms.) Includes 1 second readout update interval.