SPECIFICATIONS
ELECTRICAL
Input Voltage: 100 - 240 VAC, 50 - 60 Hz
Input Current: 1A (@ 100 VAC)
Output: 60 VDC @ 0.4A
Leakage Current: less than 100 microamps
DIMENSIONS
8.2" (21 cm) (L) x 3.1" (8 cm) (W) x 5.2" (13 cm) (H)
ILLUMINATION SYSTEM
Welch Allyn HILux
Lamp
Color Temperature: 5500°K
Lamp: 60 Volts, 24 watts
Lamp Life: 300 hours average
DIMMING SYSTEM
Manual Shutter: 20% to 100%
FIBER OPTIC BUNDLE INTERFACE
Standard ACMI Connector
APPROVALS
UL 2601-1 IEC 801-2,3,4,5
CSA C22.2 NO 601.1-M90 EN 55011
IEC 601-1 Australia AS3200
IEC 601-1-2
CE
The CE mark on this product indicates it has been tested to and conforms
with the provisions noted within the 93/42/EEC Medical Device Directive.
Declarations of Conformity are held by Welch Allyn’s European
Representative.
European Regulatory Manager
Welch Allyn Ltd.,
Kells Road, Navan,
County Meath, Republic of Ireland.
Tel +353 46 28122
Fax +353 46 28536
ENVIRONMENT
Operating: 60° F (15° C) to 85° F (30° C)
Storage: -13° F (-25° C) to 140° F (60° C)
EQUIPMENT CLASSIFICATION
Class I, Type B, Continuous Operation
FUSES
T1.0A L250V: 1.0A, 250V, time lag (slow-blow),
low breaking capacity