Product Manual

RetinaVue
TM
Network Client
Quick Start Guide Topcon NW400
Technical Support available 8:00 AM 8:00 PM EST Monday Friday
technicalsupport@welchallyn.com I +1 (800) 535-6663, Option 2
Revision date: 2016-03; 723672 DIR 80020430 Ver. C
1. Power on Computer and Camera.
To avoid communication issues, verify that
the computer is on before powering on the
camera.
2. Ensure RetinaVue Software is Running.
When the computer starts up, it should
automatically launch the application. If the
RetinaVue Network Client application is not
already open, double-click the desktop
icon.
3. Begin a New Exam & Acquire Images.
(a) First, turn off the lights in the room so
that the patient’s pupils can begin to
dilate.
(b) Click the New Exam button in the
RetinaVue Network Client software.
(c) Complete all required fields. Required
fields are outlined in red.
(d) Click the NEXT button to save the
patient information. A message should
be displayed indicating the patient has
been added to the camera. Press OK.
(e) On the Topcon camera screen, select
the Patient field in the top left corner.
(f) Select the Refresh button.
(g) Select the patient from the result list and
then select the button in the bottom left
corner to return to the exam window.
(h) Use the chinrest up / down buttons on
the right of the camera screen to align
the patient’s eye within the blue lines.
The blue square on the left can be used
for minor adjustments.
IMPORTANT: The patients eye should line up
with the line on the outside of the head rest
support (patient left side). The bridge of the