Product Manual

36
SECTION 4
Rhythm Leads
Displays continuous rhythm of selected ECG leads and permits printing of selected leads. User may toggle between
selected leads, system set leads, or I, II, III, aVR, aVL, and aVF followed by V1, V2, V3, V4, V5, and V6.
NOTE: Rhythm acquisition is not stored in memory, only printed.
NOTE: See Section 3 to acquire a rhythm printout.
Bar Code Scanner
Enabling this option will allow for the use of a Mortara approved USB bar code scanner.
Average RR
Enabling this option will display an averaged RR value to appear on the report.
QTcB
Enabling this option will display a Bazett’s corrected QT value on the report along with the default linear
QTc value.
QTcF
Enabling this option will display a Fridericia corrected QT value on the report along with the default linear
QTc value.
ECG Capture
Up to 5 minutes accumulated ECG data can be acquired internally for use with the Best 10 feature. The device
automatically selects the best 10 seconds from within the 5-minute buffer.
Users can switch between BEST 10 or LAST 10 by selecting F5 (More) followed by F5 (Last) or F5 (Best)
depending on the current view.
Encryption Key
Encryption Key is a sequence of numbers used to encrypt or decrypt data and can consist of up to 16 alpha/numeric
characters. All transmissions from x50c are sent using AES 256 bit encryption. The encryption key in the config
menu allows the user to set a custom pre-shared key between the cart and ELI Link version 4.00 or later (all carts
transmitting to one ELI Link must share the key) otherwise a default hidden key will be used.
Band Mode
Use 850/1900 MHz (US) or 900/1800 MHz (EU). (Applies to ELI 150c only.)