User's Manual Part 10

168 Specifications Welch Allyn Propaq LT Vital Signs Monitor
NIBP
Single Use - OxiCliq
OxiCliq A Adult Finger, thumb or toe 70% to 100% ± 2.5%
OxiCliq P Pediatric Finger, thumb or toe 70% to 100% ± 2.5%
OxiCliq N Adult Finger 70% to 100% ± 2.5%
OxiCliq N Neonate Foot or hand 70% to 100% ± 3.5%
OxiCliq I Infant Great toe or thumb 70% to 100% ± 2.5%
Reusable
D
-YS DURA-Y, Multisite (with D-YSE, Earclip
for use with D
-YS)
Ear lobe or pinna 70% to 100% ± 3.5%
D-YS Neonate Ear lobe or pinna 70% to 100% ± 4%
D-YS Infant to Adult Ear lobe or pinna 70% to 100% ± 3%
DS
-100A DURASENSOR Adult Index finger 70% to 100% ± 3%
OXI
-A/N OXIBAND Adult Finger, thumb or toe 70% to 100% ± 3%
OXI
-A/N OXIBAND Neonatal Foot or hand 70% to 100% ± 4%
OXI
-P/I Pediatric Finger, thumb or great toe 70% to 100% ± 3%
OXI
-P/I Infant Great toe 70% to 100% ± 3%
a. Probe accuracy for saturation levels below 70% is unspecified.
b. When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased by ± 1 digit as compared to
adult usage, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood. For example,
MAX
-N accuracy on neonates is ± 3 digits, rather than ± 2 digits for adults.
c. The accuracy specification under motion conditions is ± 3.
Table 16. Specification: Pulse Oximetry (SpO
2
) Nellcor Sensor (continued)
Table 17. Specification: Noninvasive Blood Pressure (NIBP)
Characteristic Specification
Method Oscillometric
Regulatory standards Meets AAMI SP10:2002, IEC 60601
-1:1990, IEC 60601-2-30:1999(E);
EN 1060
-1:1996, EN 1060-3:1997
Control Automatic (at selected intervals), turbo, and manual
Auto intervals 1, 2, 3, 5, 10, 15, 30, and 60 minutes
Smartcuf (Available in 2006) Available when ECG is used on adult patients
Reported pressures Systolic, diastolic, and mean plus real
-time manometer pressure
Pressure resolution 1 mmHg