User's Manual Part 10

Directions for Use Specifications 173
Environmental
Physical
Caution The monitor might not meet performance specifications if it is not used
or stored within these environmental specifications.
Note
The specifications apply when the battery is installed and the battery cover is in
place.
Table 21. Specification
Characteristic Specification
Operating temperature 0° to 40° C
Shipping and storage temperature
-20° to 60° C for monitor and cradle.
See Table 25, “Specification: Monitor Battery” on page 175 for details about
long
-term storage of the internal monitor battery.
Operating altitude
-2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and storage altitude
-2,000 to 40,000 ft (-610 to 12,192 m)
Operating relative humidity 5% to 95%, noncondensing per MIL STD 810E, Procedure
1-natural
Shipping and storage relative humidity 5% to 95%, noncondensing per MIL STD 810E, Procedure
1-natural
Water resistance IPX1
Drop 2 meter onto vinyl tile over concrete per EN60601
-1
(
Cosmetic damage is not considered a failure.)
Shock 75 g, 11 ms half sine waveform, three shocks per face for a total of 18 shocks.
Vibration, random 0.02g
2
/Hz from 10 to 500 Hz, ramping down to 0.002g
2
/Hz at 2000 Hz.
Superimposed sine frequencies of 30 Hz at 2.5 g and 60, 90, and 120 Hz at 1.5
g. Operating 1 hour per axis, 3 hours per test.
Electromagnetic compatibility (EMC) EN 60601
-1-2: 2001, Emission Class A for monitor with Large Color Display
Interface, Emission Class B for monitor without Large Color Display Interface
Table 22. Specification
Characteristic Specification
Protection Classifications, all Configurations
Type of protection against electric shock—
monitor (connected to internal battery power
source or to cradle.)
Battery operation or connection to isolated cradle. Battery must be
charged in the monitor while the monitor is in the cradle.
Degree of protection against electric shock, for
parts applied to patients
See monitor labels. Type CF, defibrillator protected.